FDA Adverse Event
Malfunction
Summary report: N
EQUAL Y CONNECTOR 1/4
MDR report key: 16277046
·
Received February 1, 2023
Report
- Report Number
- 16277046
- Event Type
- Malfunction
- Date Received
- February 1, 2023
- Date of Event
- December 16, 2022
- Report Date
- January 20, 2023
- Manufacturer
- NOVOSCI CORP.
- Product Code
- KRI
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- ME
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
Description of Event or Problem · 0
NARRATIVE FROM STAFF: Y CONNECTOR NOT DRAINING. NOTED THAT IT WAS DEFECTIVE AND CHANGED OUT. IT WAS NOTED THAT THE CHEST TUBE NOT DRAINING WHEN EXAMINED NOTED THAT A PLASTIC BARRIER (DEFECT) INSIDE THE Y CONNECTOR WAS PREVENTING THE DRAINAGE FROM FLOWING. PATIENT PUT OUT 150ML WHEN Y CONNECTOR WAS CHANGED OUT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1034629 | EQUAL Y CONNECTOR 1/4 | ACCESSORY EQUIPMENT, CARDIOPULMONARY BYPASS | KRI | NOVOSCI CORP. | C320S | 303897 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 20440 DA | Male |