FDA Adverse Event Malfunction Summary report: N

PUREPOINT LASER

MDR report key: 16276363 · Received February 1, 2023

Report

Report Number
2028159-2023-00151
Event Type
Malfunction
Date Received
February 1, 2023
Date of Event
January 12, 2023
Report Date
March 20, 2023
Manufacturer
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
Product Code
HQF
UDI-DI
00380657505975
PMA / PMN Number
K062624
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE COMPANY REPRESENTATIVE CONFIRMED AND REPLICATED THE REPORTED EVENT. THE PHOTO MONITORS WERE CALIBRATED TO RESOLVE THE ISSUE. THE SYSTEM WAS TESTED AND FOUND TO MEET PRODUCT SPECIFICATIONS. A NON-CONFORMANCE BASED REVIEW OF THE SERIAL NUMBER WAS PERFORMED AND DID NOT REVEAL ANY POTENTIAL CONTRIBUTING FACTORS TO THE REPORTED COMPLAINT. A MANUFACTURING DEVICE HISTORY RECORD (DHR) REVIEW WAS PERFORMED PRIOR TO PRODUCT RELEASE TO ENSURE THAT THE PRODUCT WAS MANUFACTURED IN COMPLIANCE WITH THE DEVICE MASTER RECORD. BASED ON THE ASSESSMENT, THE PRODUCT MET RELEASE CRITERIA. A REVIEW FOR COMPLAINTS REPORTED AGAINST THIS SERIAL NUMBER WAS PERFORMED. NO SIMILAR COMPLAINTS WERE REPORTED FOR THE PRODUCT SERIAL UNDER INVESTIGATION. THE ROOT CAUSE OF THE REPORTED EVENT IS ATTRIBUTED TO PHOTO MONITOR OUT OF CALIBRATION DUE TO COMPONENT WEAR. MANUFACTURER WILL CONTINUE TO MONITOR DATA FOR EVIDENCE OF ADVERSE TRENDING AND TAKE FURTHER ACTION, AS APPROPRIATE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Additional Manufacturer Narrative · 0

INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Description of Event or Problem · 0

A CUSTOMER REPORTED THAT AN OPHTHALMIC LASER EXHIBITED LOW LASER ENERGY. PROCEDURE WAS COMPLETED. PATIENT IMPACT DETAILS WERE NOT BEEN PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
619970 PUREPOINT LASER LASER, OPHTHALMIC HQF ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER NA NA 00380657505975

Patients

Seq Age Sex Outcome Treatment
1 Unknown