FDA Adverse Event
Malfunction
Summary report: N
PENTAX
MDR report key: 16275857
·
Received February 1, 2023
Report
- Report Number
- 9610877-2023-50984
- Event Type
- Malfunction
- Date Received
- February 1, 2023
- Date of Event
- January 25, 2023
- Report Date
- February 1, 2023
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- PEA
- UDI-DI
- 04961333221141
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- RS
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 0
THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EPK-I7010-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K182004. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE IRIS DIAPHRAGM IMAGE FAILURE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE SHOCK APPLIED ON THE IRIS DIAPHRAGM. BASED ON THE TECHNICAL REPORT AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
Description of Event or Problem · 0
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. IRIS DIAPHRAGM FAILURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1402503 | PENTAX | OPTIVISTA HD VIDEO PROCESSOR (INTERNATIONAL) | PEA | HOYA CORPORATION PENTAX TOKYO OFFICE | EPK-I7010 | 04961333221141 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |