FDA Adverse Event Malfunction Summary report: N

BD BBL¿ COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD

MDR report key: 16272558 · Received January 31, 2023

Report

Report Number
9680577-2023-00002
Event Type
Malfunction
Date Received
January 31, 2023
Date of Event
January 4, 2023
Report Date
February 23, 2023
Manufacturer
BECTON DICKINSON GMBH
Product Code
JSJ
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY THIS MEMO IS TO SUMMARIZE FINDINGS ON YOUR RECENT COMPLAINT 6892292 AGAINST COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD, CATALOG NUMBER 254007, LOT NUMBER UNKNOWN EVENT DESCRIPTION: THE CUSTOMER REPORTED A POORLY SELECTIVE MEDIA, COMPLAINT HISTORY REVIEW: THE COMPLAINT HISTORY HAS BEEN REVIEWED FOR A PERIOD OF 12 MONTHS AND SIMILAR ISSUE WERE REPORTED FOR THIS CATALOG NUMBER. HOWEVER NO TREND WAS IDENTIFIED. BATCH HISTORY RECORD (BHR) REVIEW: THE BHR COULD NOT BE REVIEWED SINCE THE RIGHT BATCH NUMBER WAS NOT COMMUNICATED. SAMPLE ANALYSIS: RETAIN SAMPLES WERE NOT REVIEWED SINCE THE RIGHT BATCH NUMBER COULD NOT BE COMMUNICATED. RETURN SAMPLES WERE NOT PROVIDED HOWEVER PICTURE ILLUSTRATING A PLATE WITH BACTERIAL GROWTH WAS PROVIDED. EVALUATION RESULTS: BASED UPON OUR INVESTIGATION, WE HAVE EXCLUDED ANY SYSTEMATIC FAILURE IN OUR MANUFACTURING PROCESS. SINCE THE RIGHT BATCH NUMBER WAS NOT COMMUNICATED NO FURTHER INVESTIGATION COULD BE DETECTED. INVESTIGATION CONCLUSION: BASED ON THE ABOVE MENTIONED EVALUATION, THE COMPLAINT CANNOT BE CONFIRMED. HOWEVER BD WILL CONTINUE MONITORING INCOMING COMPLAINT WITH SIMILAR FAILURE MODE.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. MEDICAL DEVICE LOT #: 231380 WAS REPORTED, HOWEVER, THIS IS NOT A LOT# MANUFACTURED FOR THIS PRODUCT.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD BBL¿ COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD GRAM NEGATIVE BACTERIA GROWTH AND COVERS THE GROWTH OF GRAM-POSITIVE GROWTH. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: REPORTS FOR LOT: 231380 EXPIRING 30/01/2023 AS A POORLY SELECTIVE MEDIA, GRAM-NEGATIVE BACTERIA FREQUENTLY GROW, COVERING THE GROWTH OF GRAM-POSITIVE BACTERIA, AND SOMETIMES GRAM-POSITIVE BACTERIA ARE INHIBITED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD BBL¿ COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD GRAM NEGATIVE BACTERIA GROWTH AND COVERS THE GROWTH OF GRAM-POSITIVE GROWTH. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: REPORTS FOR LOT: 231380, EXPIRING 30/01/2023 AS A POORLY SELECTIVE MEDIA, GRAM-NEGATIVE BACTERIA FREQUENTLY GROW, COVERING THE GROWTH OF GRAM-POSITIVE BACTERIA, AND SOMETIMES GRAM-POSITIVE BACTERIA ARE INHIBITED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
789391 BD BBL¿ COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD CULTURE MEDIA, SELECTIVE AND NON-DIFFERENTIAL JSJ BECTON DICKINSON GMBH

Patients

Seq Age Sex Outcome Treatment
1 Unknown