BD BBL¿ COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD
Report
- Report Number
- 9680577-2023-00002
- Event Type
- Malfunction
- Date Received
- January 31, 2023
- Date of Event
- January 4, 2023
- Report Date
- February 23, 2023
- Manufacturer
- BECTON DICKINSON GMBH
- Product Code
- JSJ
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H.6 INVESTIGATION SUMMARY THIS MEMO IS TO SUMMARIZE FINDINGS ON YOUR RECENT COMPLAINT 6892292 AGAINST COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD, CATALOG NUMBER 254007, LOT NUMBER UNKNOWN EVENT DESCRIPTION: THE CUSTOMER REPORTED A POORLY SELECTIVE MEDIA, COMPLAINT HISTORY REVIEW: THE COMPLAINT HISTORY HAS BEEN REVIEWED FOR A PERIOD OF 12 MONTHS AND SIMILAR ISSUE WERE REPORTED FOR THIS CATALOG NUMBER. HOWEVER NO TREND WAS IDENTIFIED. BATCH HISTORY RECORD (BHR) REVIEW: THE BHR COULD NOT BE REVIEWED SINCE THE RIGHT BATCH NUMBER WAS NOT COMMUNICATED. SAMPLE ANALYSIS: RETAIN SAMPLES WERE NOT REVIEWED SINCE THE RIGHT BATCH NUMBER COULD NOT BE COMMUNICATED. RETURN SAMPLES WERE NOT PROVIDED HOWEVER PICTURE ILLUSTRATING A PLATE WITH BACTERIAL GROWTH WAS PROVIDED. EVALUATION RESULTS: BASED UPON OUR INVESTIGATION, WE HAVE EXCLUDED ANY SYSTEMATIC FAILURE IN OUR MANUFACTURING PROCESS. SINCE THE RIGHT BATCH NUMBER WAS NOT COMMUNICATED NO FURTHER INVESTIGATION COULD BE DETECTED. INVESTIGATION CONCLUSION: BASED ON THE ABOVE MENTIONED EVALUATION, THE COMPLAINT CANNOT BE CONFIRMED. HOWEVER BD WILL CONTINUE MONITORING INCOMING COMPLAINT WITH SIMILAR FAILURE MODE.
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. MEDICAL DEVICE LOT #: 231380 WAS REPORTED, HOWEVER, THIS IS NOT A LOT# MANUFACTURED FOR THIS PRODUCT.
IT WAS REPORTED THAT BD BBL¿ COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD GRAM NEGATIVE BACTERIA GROWTH AND COVERS THE GROWTH OF GRAM-POSITIVE GROWTH. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: REPORTS FOR LOT: 231380 EXPIRING 30/01/2023 AS A POORLY SELECTIVE MEDIA, GRAM-NEGATIVE BACTERIA FREQUENTLY GROW, COVERING THE GROWTH OF GRAM-POSITIVE BACTERIA, AND SOMETIMES GRAM-POSITIVE BACTERIA ARE INHIBITED.
IT WAS REPORTED THAT BD BBL¿ COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD GRAM NEGATIVE BACTERIA GROWTH AND COVERS THE GROWTH OF GRAM-POSITIVE GROWTH. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: REPORTS FOR LOT: 231380, EXPIRING 30/01/2023 AS A POORLY SELECTIVE MEDIA, GRAM-NEGATIVE BACTERIA FREQUENTLY GROW, COVERING THE GROWTH OF GRAM-POSITIVE BACTERIA, AND SOMETIMES GRAM-POSITIVE BACTERIA ARE INHIBITED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 789391 | BD BBL¿ COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD | CULTURE MEDIA, SELECTIVE AND NON-DIFFERENTIAL | JSJ | BECTON DICKINSON GMBH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |