ACUJECT VARIABLE INJECTION NEEDLE

MDR report key: 1627143 · Received March 4, 2010

Report

Report Number
1037905-2010-00100
Event Type
Malfunction
Date Received
March 4, 2010
Date of Event
January 20, 2010
Report Date
February 3, 2010
Manufacturer
COOK ENDOSCOPY
Product Code
FBK
PMA / PMN Number
K941305
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION: OUR EVALUATION OF THE RETURNED DEVICE CONFIRMED THE REPORT OF NEEDLE EXTENSION DIFFICULTIES. THE NEEDLE HAS PENETRATED THE INNER WALL OF THE OUTER CATHETER, CAUSING NEEDLE EXTENSION DIFFICULTIES. A SMALL DAMAGED AREA IN THE OUTER CATHETER WHERE THE NEEDLE PENETRATION OCCURRED WAS DETECTED THROUGH A VISUAL EXAMINATION. THE SMALL DAMAGED AREA IS LOCATED IN THE BLACK SECTION OF THE OUTER CATHETER NEAR THE DISTAL END. A DISCREPANCY OR ANOMALY SPECIFIC TO THE PRODUCT THAT COULD HAVE CAUSED THIS OCCURRENCE WAS NOT OBSERVED DURING OUR LABORATORY ANALYSIS. AFTER A REVIEW OF THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED, WE CAN REPORT NO DISCREPANCIES OR ANOMALIES WERE OBSERVED. WE WERE UNABLE TO CONDUCT A SAMPLE TEST FROM THIS LOT BECAUSE NONE OF THE PRODUCT FROM THIS LOT REMAINED IN FINISHED GOODS INVENTORY. A REVIEW OF THE TWELVE MONTH COMPLAINT HISTORY FOR THIS PRODUCT FAMILY WAS CONDUCTED. BASED ON THIS REVIEW, THE LIKELIHOOD OF THIS TYPE OF REPORT IS RARE. CONCLUSIONS: NEEDLE PENETRATION OF THE OUTER CATHETER CAN OCCUR IF NEEDLE EXTENSION IS ATTEMPTED WITH THE CATHETER IN A COILED OR CURVED POSITION. THE INSTRUCTIONS FOR USE DIRECT THE USER TO UNCOIL THE CATHETER AND STRAIGHTEN COMPLETELY. THE INSTRUCTIONS FOR USE ALSO CAUTION THE USER THAT EXTENDING THE NEEDLE WHILE THE CATHETER IS COILED MAY RESULT IN DAMAGE TO THE PERFORMANCE CHARACTERISTICS OF THE DEVICE. PRIOR TO DISTRIBUTION, ALL ACUJECT VARIABLE INJECTION NEEDLES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TEST TO ENSURE APPROPRIATE NEEDLE EXTENSION. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A CORRECTIVE ACTION HAS BEEN INITIATED IN AN EFFORT TO REDUCE OCCURRENCES OF THIS NATURE. THIS PRODUCT WAS MANUFACTURED PRIOR TO IMPLEMENTATION OF THIS CORRECTIVE ACTION. THE LIKELIHOOD OF THIS TYPE OF OCCURRENCE IS RARE. CUSTOMER QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS.

Description of Event or Problem · 1

DURING A COLONOSCOPY, THE PHYSICIAN USED A COOK ENDOSCOPY ACUJECT VARIABLE INJECTION NEEDLE. THE NEEDLE WAS TO BE USED FOR SALINE INJECTION OF A POLYP. THE NEEDLE WOULD NOT EXTEND BEYOND THE DISTAL END OF THE OUTER CATHETER. THE USER REPORTEDLY "TRIED TO WIGGLE IT FREE" AND THE NEEDLE PENETRATED THE SIDE OF OUTER SHEATH. THE INJECTION NEEDLE WAS REMOVED AND ANOTHER NEEDLE WAS USED TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL MEDICAL PROCEDURES DUE TO THIS OCCURRENCE. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACUJECT VARIABLE INJECTION NEEDLE FBK, ENDOSCOPIC INJECTION NEEDLE, GASTROENTEROLOGY-UROLOGY FBK COOK ENDOSCOPY W2767567

Patients

Seq Age Sex Outcome Treatment
1 UNK PROCEDURE)| FUJINON 450 SERIES ENDOSCOPE (USED DURING THIS