CARDIOSAVE HYBRID, TYPE B PLUG
Report
- Report Number
- 2249723-2023-00707
- Event Type
- Malfunction
- Date Received
- January 30, 2023
- Date of Event
- April 9, 2021
- Report Date
- January 12, 2023
- Manufacturer
- DATASCOPE CORP. - MAHWAH
- Product Code
- DSP
- UDI-DI
- 10607567108391
- PMA / PMN Number
- K112372
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN
- Reporter Occupation
- 003
Narratives
THE GETINGE FIELD SERVICE ENGINEER (FSE) THAT ENCOUNTERED THE ISSUE REPLACED THE UPPER PANEL ASSEMBLY (0997-00-0644) AND LABELS (0334-00-1811 AND 0334-00-1813) COMPLETED THE PM WITH FULL CALIBRATION, FUNCTIONAL AND SAFETY CHECKS TO MEET FACTORY SPECIFICATIONS. UNIT PASSED ALL CALIBRATION, FUNCTIONAL AND SAFETY TEST PER FACTORY SPECIFICATIONS. THE IABP WAS THEN RELEASED TO THE CUSTOMER AND CLEARED FOR CLINICAL SERVICE. THIS REPORT IS BEING SUBMITTED AS THE RESULT OF A RETROSPECTIVE REVIEW CONDUCTED IN CAPA 584165.
IT WAS REPORTED THAT DURING PREVENTIVE MAINTENANCE (PM) PERFORMED BY A GETINGE FIELD SERVICE ENGINEER (FSE), THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) HAD A BROKEN FIBER OPTIC DOOR THAT WOULD NOT RETRACT PROPERLY. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 15036 | CARDIOSAVE HYBRID, TYPE B PLUG | SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL | DSP | DATASCOPE CORP. - MAHWAH | 0998-00-0800-53 | 10607567108391 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | N/A. |