FDA Adverse Event Malfunction Summary report: N

CARDIOSAVE HYBRID, TYPE B PLUG

MDR report key: 16264565 · Received January 30, 2023

Report

Report Number
2249723-2023-00708
Event Type
Malfunction
Date Received
January 30, 2023
Date of Event
April 8, 2021
Report Date
January 12, 2023
Manufacturer
DATASCOPE CORP. - MAHWAH
Product Code
DSP
UDI-DI
10607567108391
PMA / PMN Number
K112372
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ND
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE GETINGE FIELD SERVICE ENGINEER (FSE) THAT ENCOUNTERED THE ISSUE REPLACED THE UPPER DISPLAY ASSEMBLY (0160-00-0127) AND USES LABELS (0334-00-1810-01 AND 0334-00-1809) AND FRAME BRACKET (0406-00-0894). THE FSE COMPLETED THE PM WITH FULL CALIBRATION, FUNCTIONAL AND SAFETY CHECKS TO MEET FACTORY SPECIFICATIONS. UNIT PASSED ALL CALIBRATION, FUNCTIONAL AND SAFETY TEST PER FACTORY SPECIFICATIONS. THE IABP WAS THEN RELEASED TO THE CUSTOMER AND CLEARED FOR CLINICAL SERVICE. THIS REPORT IS BEING SUBMITTED AS THE RESULT OF A RETROSPECTIVE REVIEW CONDUCTED IN CAPA 584165.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING PREVENTIVE MAINTENANCE (PM) PERFORMED BY A GETINGE FIELD SERVICE ENGINEER (FSE), THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) HAD A VERTICAL LINE OF RED PIXELS ON THE DISPLAY. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
15028 CARDIOSAVE HYBRID, TYPE B PLUG SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL DSP DATASCOPE CORP. - MAHWAH 0998-00-0800-53 10607567108391

Patients

Seq Age Sex Outcome Treatment
1 Unknown