BD SLIP TIP SYRINGE WITH BD PRECISIONGLIDE¿ NEEDLE
Report
- Report Number
- 8041187-2023-00036
- Event Type
- Malfunction
- Date Received
- January 28, 2023
- Date of Event
- January 10, 2023
- Report Date
- March 10, 2023
- Manufacturer
- BECTON DICKINSON MEDICAL (SINGAPORE)
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H6: INVESTIGATION SUMMARY THE BLACK DUST FOREIGN MATTER INSIDE THE PACKAGE AND SYRINGE COULD BE DUE TO PEST INFESTATION. THE FOREIGN MATTER COULD HAVE BEEN BROUGHT INTO THE PACKAGE WHEN THE PEST BIT THE BOTTOM WEB TO ENTER THE PACKAGING. THE PRODUCT COULD HAVE BEEN INFESTED WITH PEST DUE TO UNCONTROLLED STORAGE CONDITIONS BEFORE DISTRIBUTED TO THE CUSTOMERS¿ PREMISES. THE NON-CONFORMANCE COULD HAVE OCCURRED OUT OF THE MANUFACTURING FACILITY. THE TEAM HAD ALSO REVIEWED THE PEST CONTROLS RECORDS AND NO ABNORMALITIES WERE DETECTED AT THE 1ML MANUFACTURING LINE. HENCE, WE ARE UNABLE TO IDENTIFY THE ROOT CAUSE OF THE REPORTED NON-CONFORMANCE. DHR FOR SYRINGE WAS PERFORMED FOR THE BATCH 2140745 H3 OTHER TEXT : SEE H10
A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT THE BD SLIP TIP SYRINGE WITH BD PRECISIONGLIDE¿ NEEDLE HAD FOREIGN MATTER IN THE BARREL. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WHILE OPENING NEW SYRINGE NURSE OBSERVED FOREIGN MATERIAL INSIDE SYRINGE BARREL."
IT WAS REPORTED THAT THE BD SLIP TIP SYRINGE WITH BD PRECISIONGLIDE¿ NEEDLE HAD FOREIGN MATTER IN THE BARREL. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WHILE OPENING NEW SYRINGE NURSE OBSERVED FOREIGN MATERIAL INSIDE SYRINGE BARREL."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 380809 | BD SLIP TIP SYRINGE WITH BD PRECISIONGLIDE¿ NEEDLE | PISTON SYRINGE | FMF | BECTON DICKINSON MEDICAL (SINGAPORE) | 2140745 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |