FDA Adverse Event Malfunction Summary report: N

BD SLIP TIP SYRINGE WITH BD PRECISIONGLIDE¿ NEEDLE

MDR report key: 16261485 · Received January 28, 2023

Report

Report Number
8041187-2023-00036
Event Type
Malfunction
Date Received
January 28, 2023
Date of Event
January 10, 2023
Report Date
March 10, 2023
Manufacturer
BECTON DICKINSON MEDICAL (SINGAPORE)
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY THE BLACK DUST FOREIGN MATTER INSIDE THE PACKAGE AND SYRINGE COULD BE DUE TO PEST INFESTATION. THE FOREIGN MATTER COULD HAVE BEEN BROUGHT INTO THE PACKAGE WHEN THE PEST BIT THE BOTTOM WEB TO ENTER THE PACKAGING. THE PRODUCT COULD HAVE BEEN INFESTED WITH PEST DUE TO UNCONTROLLED STORAGE CONDITIONS BEFORE DISTRIBUTED TO THE CUSTOMERS¿ PREMISES. THE NON-CONFORMANCE COULD HAVE OCCURRED OUT OF THE MANUFACTURING FACILITY. THE TEAM HAD ALSO REVIEWED THE PEST CONTROLS RECORDS AND NO ABNORMALITIES WERE DETECTED AT THE 1ML MANUFACTURING LINE. HENCE, WE ARE UNABLE TO IDENTIFY THE ROOT CAUSE OF THE REPORTED NON-CONFORMANCE. DHR FOR SYRINGE WAS PERFORMED FOR THE BATCH 2140745 H3 OTHER TEXT : SEE H10

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD SLIP TIP SYRINGE WITH BD PRECISIONGLIDE¿ NEEDLE HAD FOREIGN MATTER IN THE BARREL. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WHILE OPENING NEW SYRINGE NURSE OBSERVED FOREIGN MATERIAL INSIDE SYRINGE BARREL."

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD SLIP TIP SYRINGE WITH BD PRECISIONGLIDE¿ NEEDLE HAD FOREIGN MATTER IN THE BARREL. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WHILE OPENING NEW SYRINGE NURSE OBSERVED FOREIGN MATERIAL INSIDE SYRINGE BARREL."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
380809 BD SLIP TIP SYRINGE WITH BD PRECISIONGLIDE¿ NEEDLE PISTON SYRINGE FMF BECTON DICKINSON MEDICAL (SINGAPORE) 2140745

Patients

Seq Age Sex Outcome Treatment
1 Unknown