BD PLASTIPAK¿ SYRINGES
Report
- Report Number
- 3003152976-2023-00009
- Event Type
- Malfunction
- Date Received
- January 28, 2023
- Date of Event
- December 14, 2022
- Report Date
- March 13, 2023
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMF
- UDI-DI
- 00382903008650
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
H6: INVESTIGATION SUMMARY NO PHOTOS OR PHYSICAL SAMPLES THAT DISPLAY THE REPORTED CONDITION HAVE BEEN PROVIDED FOR EVALUATION. A DEVICE HISTORY REVIEW WAS PERFORMED FOR REPORTED LOT 2111021, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. TEN RETAINED SAMPLES OF LOT 2111021 WERE USED FOR ADDITIONAL EVALUATION. THE PRODUCTS WERE VISUALLY INSPECTED, NO FOREIGN MATTER OR OTHER LIQUID WAS OBSERVED INSIDE ANY OF THE SYRINGES. IT IS POSSIBLE THE LIQUID THAT WAS DESCRIBED TO BE OBSERVED IN THE SYRINGE WAS SILICONE. LUBRICANT IS EMPLOYED DURING THE SYRINGE ASSEMBLY PROCESS TO LUBRICATE THE CYLINDERS IN THE SILICONE STATION. THE SILICONE EMPLOYED IN THIS PRODUCT IS A MEDICAL GRADE SILICONE AUTHORIZED FOR PRODUCT USE. SILICONE CONTENT TESTS ARE PERFORMED DURING THE MANUFACTURING PROCESS OF EACH LOT NUMBER. RESULTS WERE REVIEWED FOR LOT 2111021 AND FOUND TO BE WITHIN SPECIFICATION. SILICONE CAN BE VISIBLE DUE TO ISSUES RELATED TO POOR DISTRIBUTION INSIDE THE BARREL. AS NO DEFECT WAS OBSERVED IN THE RETAINED SAMPLES AND DEVICE RECORDS DO NOT INDICATE ANY ISSUES OCCURRED DURING MANUFACTURING, A DEFINITIVE ROOT CAUSE RELATED TO OUR MANUFACTURING PROCESS CANNOT BE IDENTIFIED AT THIS TIME. H3 OTHER TEXT : SEE H10.
IT WAS REPORTED WHILE USING BD PLASTIPAK¿ SYRINGES FOREIGN MATTER WAS FOUND. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: RELEASE OF SILICONE FROM THE PLUNGER OF THE SYRINGE.
IT WAS REPORTED WHILE USING BD PLASTIPAK¿ SYRINGES FOREIGN MATTER WAS FOUND. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: RELEASE OF SILICONE FROM THE PLUNGER OF THE SYRINGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 372109 | BD PLASTIPAK¿ SYRINGES | PISTON SYRINGE | FMF | BECTON DICKINSON, S.A. | 2111021 | 00382903008650 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |