FDA Adverse Event Malfunction Summary report: N

BD PLASTIPAK¿ SYRINGES

MDR report key: 16261474 · Received January 28, 2023

Report

Report Number
3003152976-2023-00009
Event Type
Malfunction
Date Received
January 28, 2023
Date of Event
December 14, 2022
Report Date
March 13, 2023
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
UDI-DI
00382903008650
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY NO PHOTOS OR PHYSICAL SAMPLES THAT DISPLAY THE REPORTED CONDITION HAVE BEEN PROVIDED FOR EVALUATION. A DEVICE HISTORY REVIEW WAS PERFORMED FOR REPORTED LOT 2111021, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. TEN RETAINED SAMPLES OF LOT 2111021 WERE USED FOR ADDITIONAL EVALUATION. THE PRODUCTS WERE VISUALLY INSPECTED, NO FOREIGN MATTER OR OTHER LIQUID WAS OBSERVED INSIDE ANY OF THE SYRINGES. IT IS POSSIBLE THE LIQUID THAT WAS DESCRIBED TO BE OBSERVED IN THE SYRINGE WAS SILICONE. LUBRICANT IS EMPLOYED DURING THE SYRINGE ASSEMBLY PROCESS TO LUBRICATE THE CYLINDERS IN THE SILICONE STATION. THE SILICONE EMPLOYED IN THIS PRODUCT IS A MEDICAL GRADE SILICONE AUTHORIZED FOR PRODUCT USE. SILICONE CONTENT TESTS ARE PERFORMED DURING THE MANUFACTURING PROCESS OF EACH LOT NUMBER. RESULTS WERE REVIEWED FOR LOT 2111021 AND FOUND TO BE WITHIN SPECIFICATION. SILICONE CAN BE VISIBLE DUE TO ISSUES RELATED TO POOR DISTRIBUTION INSIDE THE BARREL. AS NO DEFECT WAS OBSERVED IN THE RETAINED SAMPLES AND DEVICE RECORDS DO NOT INDICATE ANY ISSUES OCCURRED DURING MANUFACTURING, A DEFINITIVE ROOT CAUSE RELATED TO OUR MANUFACTURING PROCESS CANNOT BE IDENTIFIED AT THIS TIME. H3 OTHER TEXT : SEE H10.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD PLASTIPAK¿ SYRINGES FOREIGN MATTER WAS FOUND. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: RELEASE OF SILICONE FROM THE PLUNGER OF THE SYRINGE.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD PLASTIPAK¿ SYRINGES FOREIGN MATTER WAS FOUND. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: RELEASE OF SILICONE FROM THE PLUNGER OF THE SYRINGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
372109 BD PLASTIPAK¿ SYRINGES PISTON SYRINGE FMF BECTON DICKINSON, S.A. 2111021 00382903008650

Patients

Seq Age Sex Outcome Treatment
1 Unknown