FDA Adverse Event Malfunction Summary report: N

BD DISCARDIT¿ II SYRINGE

MDR report key: 16261203 · Received January 28, 2023

Report

Report Number
3002682307-2023-00005
Event Type
Malfunction
Date Received
January 28, 2023
Date of Event
December 22, 2022
Report Date
February 13, 2023
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 309110 AND LOT NUMBER 2211205. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS SAMPLES WERE UNAVAILABLE FOR RETURN, TWENTY RETAINED SAMPLES WERE OBTAINED FROM THE MANUFACTURING FACILITY FOR EVALUATION; HOWEVER, THE RETAINED SAMPLES DID NOT SHOW ANY SIGNS OF LEAKAGE. IT HAS BEEN CONCLUDED THAT THE REPORTED INCIDENT MAY HAVE RESULTED FROM DAMAGE TO PLUNGER COMPONENT.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT MEDICATION LEAKED FROM THE BACK OF THE BD DISCARDIT¿ II SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM GERMAN: "THE MEDICATION RUNS OUT OF THE BACK OF THE CONE."

Description of Event or Problem · 0

IT WAS REPORTED THAT MEDICATION LEAKED FROM THE BACK OF THE BD DISCARDIT¿ II SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM GERMAN: "THE MEDICATION RUNS OUT OF THE BACK OF THE CONE.".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
372081 BD DISCARDIT¿ II SYRINGE PISTON SYRINGE FMF BECTON DICKINSON, S.A. 2211205

Patients

Seq Age Sex Outcome Treatment
1 Unknown