BD DISCARDIT¿ II SYRINGE
Report
- Report Number
- 3002682307-2023-00005
- Event Type
- Malfunction
- Date Received
- January 28, 2023
- Date of Event
- December 22, 2022
- Report Date
- February 13, 2023
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H6: INVESTIGATION SUMMARY A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 309110 AND LOT NUMBER 2211205. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS SAMPLES WERE UNAVAILABLE FOR RETURN, TWENTY RETAINED SAMPLES WERE OBTAINED FROM THE MANUFACTURING FACILITY FOR EVALUATION; HOWEVER, THE RETAINED SAMPLES DID NOT SHOW ANY SIGNS OF LEAKAGE. IT HAS BEEN CONCLUDED THAT THE REPORTED INCIDENT MAY HAVE RESULTED FROM DAMAGE TO PLUNGER COMPONENT.
A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT MEDICATION LEAKED FROM THE BACK OF THE BD DISCARDIT¿ II SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM GERMAN: "THE MEDICATION RUNS OUT OF THE BACK OF THE CONE."
IT WAS REPORTED THAT MEDICATION LEAKED FROM THE BACK OF THE BD DISCARDIT¿ II SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM GERMAN: "THE MEDICATION RUNS OUT OF THE BACK OF THE CONE.".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 372081 | BD DISCARDIT¿ II SYRINGE | PISTON SYRINGE | FMF | BECTON DICKINSON, S.A. | 2211205 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |