FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK¿ 1-ML SYRINGE

MDR report key: 16261154 · Received January 28, 2023

Report

Report Number
1213809-2023-00043
Event Type
Malfunction
Date Received
January 28, 2023
Date of Event
January 9, 2023
Report Date
February 9, 2023
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
00382903096282
PMA / PMN Number
K941562
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY: ONE PHOTO WAS PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION, IT WAS OBSERVED THAT SEVERAL DUST-LIKE FIBERS PRESENT ON AND IN THE TIP OF THE SYRINGE. POTENTIAL ROOT CAUSE FOR THE LOOSE FOREIGN MATTER IN THE UNIT PACKAGE DEFECT IS ASSOCIATED WITH THE PACKAGING PROCESS. IT IS POSSIBLE THAT PARTICULATE/FIBER ACCUMULATION DUE TO EXCESS STATIC ELECTRICITY WAS PRESENT ON THE TOP WEB PACKAGING WHICH RESULTED IN THE FOREIGN MATTER BEING SEALED IN THE UNIT PACKAGE. SINCE MANUFACTURE OF THIS BATCH A TOP WEB DEIONIZER HAS BEEN INSTALLED ON THIS LINE TO AID IN REDUCTION OF STATIC BUILDUP ON PACKAGING MATERIAL. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 0043411. A REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD LUER-LOK¿ 1-ML SYRINGE FOREIGN MATTER WAS FOUND. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: LUER-LOK 1ML SYRINGE HAD UNKNOWN SUBSTANCE, POSSIBLY CARDBOARD ATTACHED TO THE SYRINGE TIP. 10-JAN-2023 - PICTURE REVIEW BY RCC, BROWNISH FOREIGN MATTER OBSERVED INSIDE THE SEALED PACKAGING.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD LUER-LOK¿ 1-ML SYRINGE FOREIGN MATTER WAS FOUND. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: LUER-LOK 1ML SYRINGE HAD UNKNOWN SUBSTANCE, POSSIBLY CARDBOARD ATTACHED TO THE SYRINGE TIP. (B)(6) 2023 - PICTURE REVIEW BY RCC, BROWNISH FOREIGN MATTER OBSERVED INSIDE THE SEALED PACKAGING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
846282 BD LUER-LOK¿ 1-ML SYRINGE PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 0043411 00382903096282

Patients

Seq Age Sex Outcome Treatment
1 Unknown