OMNIPOD INSULIN PUMP
Report
- Report Number
- 3004464228-2010-01077
- Event Type
- Other
- Date Received
- March 4, 2010
- Date of Event
- February 8, 2010
- Report Date
- February 8, 2010
- Manufacturer
- INSULET CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042792
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE PRODUCT WILL NOT BE RETURNED TO THE MFR FOR EVAL. WE ARE UNABLE TO CONFIRM ANY PRODUCT MALFUNCTION. NO CONCLUSION CAN BE REACHED AT THIS TIME. THE CUSTOMER STATED THAT A KINK WAS NOTED IN THE CANNULA. THERE WAS, HOWEVER, NO REPORT OF AN OCCLUSION ALARM. THE POD WOULD HAVE INITIATED AN OCCLUSION ALARM IF THE FLOW OF INSULIN WAS RESTRICTED/PREVENTED BY THE KINK AND RESULTED IN BACK PRESSURE. THE PRODUCT USER GUIDE INSTRUCTS THE USER TO "CHECK INFUSION SITE FREQUENTLY FOR PROPER CANNULA PLACEMENT". IT ALSO SUGGESTS THAT THE USER SHOULD CHECK THEIR BLOOD GLUCOSE LEVELS FREQUENTLY IN ORDER TO NOTICE AND REACT QUICKLY AND APPROPRIATELY TO ANY ISSUES. THE PT REMEDIED THE SITUATION BY REMOVING AND REPLACING THE DEVICE PER USER INSTRUCTIONS.
THE CUSTOMER REPORTED THAT HE EXPERIENCED "SOME HIGH BGS WITH THE POD HE WAS WEARING" (277 MG/DL), WHICH HE ATTRIBUTES "TO THE SITE" SINCE HE IS "LEAN AND MUSCULAR." WHEN THE POD WAS REMOVED, HOWEVER, A KINK WAS SEEN IN THE CANNULA; NO POD ALARM WAS INITIATED. THE POD WILL NOT BE RETURNED FOR EVAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OMNIPOD INSULIN PUMP | PUMP, INFUSION, INSULIN | LZG | INSULET CORPORATION | 11200 | L30246 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR | Other |