BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER
Report
- Report Number
- 1710034-2023-00042
- Event Type
- Malfunction
- Date Received
- January 27, 2023
- Date of Event
- January 9, 2023
- Report Date
- March 7, 2023
- Manufacturer
- BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
- Product Code
- FOZ
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL: YES. D10: RETURNED TO MANUFACTURER ON: 30-JAN-2023. H6: INVESTIGATION SUMMARY: OUR QUALITY ENGINEER INSPECTED THE SAMPLE AND PHOTOGRAPH SUBMITTED FOR EVALUATION. BD RECEIVED ONE SEALED 22G X 1.00IN. INSYTE AUTOGUARD BC PRO UNIT FROM LOT NUMBER 2241107. ADDITIONALLY, ONE PHOTO IS PROVIDED. THE REPORT OF A NEEDLE RETRACTION FAILURE COULD NOT BE CONFIRMED FROM THE PHOTO OR SEALED REPRESENTATIVE 22G INSYTE AUTOGUARD UNIT THAT WERE PROVIDED FOR INVESTIGATION. A VISUAL INSPECTION AND FUNCTIONAL TEST REVEALED NOTHING REMARKABLE. THE NEEDLE RETRACTED SUCCESSFULLY WHEN THE SAFETY MECHANISM WAS ACTIVATED. THE RETURNED UNIT PROVIDED FOR EVALUATION MET AND PERFORMED PER THE REQUIRED MANUFACTURING SPECIFICATIONS. THERE WAS NO PHYSICAL/MECHANICAL EVIDENCE TO CONFIRM AND SUPPORT A MANUFACTURING PROCESS RELATED ISSUE FOR THE FAILURE STATED IN YOUR REPORT. A DEVICE HISTORY RECORD REVIEW SHOWED NO NON-CONFORMANCES ASSOCIATED WITH THIS ISSUE DURING THE PRODUCTION OF THIS BATCH. HOWEVER, A TREND HAS BEEN IDENTIFIED FOR THE REPORTED FAILURE AND CORRECTIVE ACTIONS HAVE BEEN INITIATED TO INVESTIGATE THIS TYPE OF INCIDENT AND IDENTITY THE ROOT CAUSE.
A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT THE NEEDLE IN THE BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER WOULD NOT RETRACT WHEN THE BUTTON WAS PRESSED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WHEN PUSH BUTTON, NEEDLE DOES NOT RETRACT".
IT WAS REPORTED THAT THE NEEDLE IN THE BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER WOULD NOT RETRACT WHEN THE BUTTON WAS PRESSED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WHEN PUSH BUTTON, NEEDLE DOES NOT RETRACT."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 470150 | BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER | INTRAVASCULAR CATHETER | FOZ | BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. | 2241107 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |