FDA Adverse Event Malfunction Summary report: N

BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER

MDR report key: 16251534 · Received January 27, 2023

Report

Report Number
1710034-2023-00042
Event Type
Malfunction
Date Received
January 27, 2023
Date of Event
January 9, 2023
Report Date
March 7, 2023
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL: YES. D10: RETURNED TO MANUFACTURER ON: 30-JAN-2023. H6: INVESTIGATION SUMMARY: OUR QUALITY ENGINEER INSPECTED THE SAMPLE AND PHOTOGRAPH SUBMITTED FOR EVALUATION. BD RECEIVED ONE SEALED 22G X 1.00IN. INSYTE AUTOGUARD BC PRO UNIT FROM LOT NUMBER 2241107. ADDITIONALLY, ONE PHOTO IS PROVIDED. THE REPORT OF A NEEDLE RETRACTION FAILURE COULD NOT BE CONFIRMED FROM THE PHOTO OR SEALED REPRESENTATIVE 22G INSYTE AUTOGUARD UNIT THAT WERE PROVIDED FOR INVESTIGATION. A VISUAL INSPECTION AND FUNCTIONAL TEST REVEALED NOTHING REMARKABLE. THE NEEDLE RETRACTED SUCCESSFULLY WHEN THE SAFETY MECHANISM WAS ACTIVATED. THE RETURNED UNIT PROVIDED FOR EVALUATION MET AND PERFORMED PER THE REQUIRED MANUFACTURING SPECIFICATIONS. THERE WAS NO PHYSICAL/MECHANICAL EVIDENCE TO CONFIRM AND SUPPORT A MANUFACTURING PROCESS RELATED ISSUE FOR THE FAILURE STATED IN YOUR REPORT. A DEVICE HISTORY RECORD REVIEW SHOWED NO NON-CONFORMANCES ASSOCIATED WITH THIS ISSUE DURING THE PRODUCTION OF THIS BATCH. HOWEVER, A TREND HAS BEEN IDENTIFIED FOR THE REPORTED FAILURE AND CORRECTIVE ACTIONS HAVE BEEN INITIATED TO INVESTIGATE THIS TYPE OF INCIDENT AND IDENTITY THE ROOT CAUSE.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE NEEDLE IN THE BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER WOULD NOT RETRACT WHEN THE BUTTON WAS PRESSED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WHEN PUSH BUTTON, NEEDLE DOES NOT RETRACT".

Description of Event or Problem · 0

IT WAS REPORTED THAT THE NEEDLE IN THE BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER WOULD NOT RETRACT WHEN THE BUTTON WAS PRESSED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WHEN PUSH BUTTON, NEEDLE DOES NOT RETRACT."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
470150 BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER INTRAVASCULAR CATHETER FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 2241107

Patients

Seq Age Sex Outcome Treatment
1 Unknown