FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM IQ INFUSION PUMP
MDR report key: 16247714
·
Received January 26, 2023
Report
- Report Number
- 1314492-2023-00109
- Event Type
- Malfunction
- Date Received
- January 26, 2023
- Report Date
- January 26, 2023
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- PMA / PMN Number
- K222048
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- NURSE
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
THE DEVICE WAS NOT RETURNED AND THE SERIAL NUMBER IS UNKNOWN; THEREFORE, A DEVICE ANALYSIS COULD NOT BE COMPLETED. THE REPORTED PROBLEM "NO DRIPS WERE FALLING" MAY POTENTIALLY BE RELATED TO FA -2021-056, WHICH IS CURRENTLY ONGOING AND IS ASSOCIATED WITH REINFORCING PROPER INTRAVENOUS LINE SETUP AND DETECTION OF UPSTREAM OCCLUSIONS FOR SPECTRUM PUMPS. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 0
IT WAS REPORTED THAT A SPECTRUM IQ PUMP HAD NO DRIPS FALLING AND BLOOD BACKING UP SEVERAL INCHES IN THE INTRAVENOUS TUBING DURING LASIX THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 785332 | SPECTRUM IQ INFUSION PUMP | PUMP, INFUSION | FRN | BAXTER HEALTHCARE CORPORATION | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | LASIX |