TRANSCAM
Report
- Report Number
- 2916556-2010-00001
- Event Type
- Other
- Date Received
- March 5, 2010
- Report Date
- February 5, 2010
- Manufacturer
- PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
- Product Code
- KPS
- PMA / PMN Number
- 921296
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- NUCLEAR MEDICINE TECHNOLOGIST
Narratives
IT WAS DETERMINED THAT THE MOST LIKELY ROOT CAUSE OF THIS EVENT WAS THAT DETECTOR MOUNTING BOLTS LOOSENED OVER TIME DUE TO EXCESSIVE VIBRATION DURING IN-HOSPITAL TRANSPORTATION. THESE LOOSE BOLTS DID NOT MAINTAIN THE DETECTOR/DETECTOR ARM INTERFACE PROPERLY WHICH CAUSED EXCESSIVE STRAIN/WEAR TO THE BOLTS AND THEY SUBSEQUENTLY FAILED ALLOWING THE DETECTOR TO FALL TO THE FLOOR. PICTURES OF THE DEVICE AND THE ACTUAL DETECTOR MOUNTING BOLTS WERE RETURNED FOR EVALUATION. A VISUAL EXAMINATION OF THE BOLTS INDICATE WEAR CONSISTENT WITH A LOOSE INTERFACE. FURTHER, IT IS POSSIBLE THAT LOOSENING OF THE BOLTS WAS ACCELERATED BY LACK OF LOCTITE ON THE BOLTS. UPON REVIEW OF THE INSTRUCTIONS FOR USE (IFU) CUSTOMERS ARE WARNED THAT DURING TRANSPORTATION THE COLUMN MUST BE IN ITS LOWEST POSITION. ALL POWER CABLES MUST BE ROLLED ONTO THE CABLE ROLL. ADDITIONALLY, THE IFU STATES THE FOLLOWING: ENTER ELEVATORS WITH THE GANTRY FORWARD AND PULL THE TRANSCAM BACKWARD TO EXIT. IF THERE IS A SAG IN THE ELEVATOR BELOW THE LEVEL OF THE FLOOR, THE LARGER WHEELS WILL MAKE EXITING MUCH EASIER. AS SPECIFIED IN MFG DOCUMENTS, (TRANSCAM ASSEMBLY MANUAL - NOVEMBER 1998) LOCTITE IS APPLIED TO THE BOLTS THAT MOUNT THE DETECTOR TO THE DETECTOR ARM. THE PHILIPS' SERVICE RECORD FROM (B)(4) 2004, STATED THAT BOLTS WERE FOUND LOOSE ON ONE SIDE OF THE DETECTOR. NEW BOLTS WERE SUBSEQUENTLY INSTALLED. WHILE THERE WAS NO DIRECT MENTION OF LOCTITE (RE)APPLICATION IN THE SERVICE RECORD, THE SERVICE MANUAL STATES TO "ALWAYS ADD THREAD-LOCKING FLUID TO ALL BOLTS." (NOTE: VISUAL EXAMINATION OF THE BOLTS WAS INCONCLUSIVE WITH RESPECT TO LOCTITE RESIDUE.) THE SYSTEM HAS BEEN SERVICED BY THIRD-PARTY VENDOR (B)(4) FOR THE PAST 5 YEARS. A REVIEW OF (B)(4) SERVICE RECORDS INDICATE THAT ROUTINE PREVENTIVE MAINTENANCE (PM) HAS BEEN COMPLETED SINCE 2007 WITH THE LAST PM BEING COMPLETED ON (B)(4) 2009. THERE ARE NO REPORTED ISSUES WITHIN THESE SERVICE RECORDS INDICATING A LOOSE DETECTOR OR BOLTS. A (B)(4) SERVICE REPORT, DATED AFTER THIS EVENT ((B)(4) 2010), STATES THAT A NEW DETECTOR WAS INSTALLED WITH NEW BOLTS AND LOCTITE AND THE SYSTEM RETURNED TO SERVICE. IT IS OUR OPINION THAT A LOOSE DETECTOR AND/OR LOOSE DETECTOR BOLTS SHOULD HAVE BEEN DETECTED (AND CORRECTED) DURING THE LAST PM. INTERNAL CROSS REFERENCE: (B)(4).
A CUSTOMER REPORTED THAT THE DETECTOR FELL OFF THE GANTRY AND ON TO THE FLOOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TRANSCAM | TRANSCAM | KPS | PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. | 5200-2338 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |