FDA Adverse Event Malfunction Summary report: N

CARDIOSAVE HYBRID TYPE B PLUG

MDR report key: 16246400 · Received January 26, 2023

Report

Report Number
2249723-2023-00490
Event Type
Malfunction
Date Received
January 26, 2023
Date of Event
August 24, 2021
Report Date
February 22, 2024
Manufacturer
DATASCOPE CORP. - MAHWAH
Product Code
DSP
UDI-DI
10607567108391
PMA / PMN Number
K112372
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

UPDATED FIELDS - (SITE COUNTRY), (TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS). A GETINGE FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO INVESTIGATE. THE FSE EVALUATED THE IABP UNIT AND CONFIRMED THAT THE HINGES WERE BENT. THE FSE RESOLVED THE ISSUE BY REPLACING THE BOTH HINGES (0105-00-0138-02) (0105-00-0138-01) AND COVERS (0380-00-0561) (0380-00-0561). THE FSE PERFORMED FUNCTIONAL AND SAFETY CHECKS TO MEET FACTORY SPECIFICATIONS. THE IABP WAS THEN RELEASED TO THE CUSTOMER AND CLEARED FOR CLINICAL SERVICE.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED AS THE RESULT OF A RETROSPECTIVE REVIEW CONDUCTED IN CAPA (B)(4). A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETED.

Description of Event or Problem · 0

N/A.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A ROUTINE CHECK, THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) SCREEN HINGES WERE BREAKING AND THE TOP DISPLAY WOULD NOT CLOSE PROPERLY. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
786250 CARDIOSAVE HYBRID TYPE B PLUG SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL DSP DATASCOPE CORP. - MAHWAH 0998-00-0800-53 10607567108391

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown