IMP,TSV,4.1MM,SBM,11.5
Report
- Report Number
- 0002023141-2023-00301
- Event Type
- Injury
- Date Received
- January 26, 2023
- Date of Event
- November 4, 2022
- Report Date
- April 13, 2023
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- UDI-DI
- 00889024019232
- PMA / PMN Number
- K072589
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
ZIMVIE COMPLAINT NUMBER CMP-0853898 THE FOLLOWING SECTIONS HAVE BEEN UPDATED: D9: DEVICE AVAILABLE FOR EVALUATION CHANGE ¿NO' TO 'YES.' H3: DEVICE EVALUATED BY MANUFACTURER: CHANGE ¿NO' TO 'YES.' PRODUCT EVALUATION: ONE IMPLANT AND MOUNT WERE RETURNED FOR INVESTIGATION. VISUAL INSPECTION OF THE AS RETURNED PRODUCT IDENTIFIED BONE DEBRIS AND A FRACTURED MOUNT. THE MOUNT WAS FRACTURED AT THE HEX. THE HEX PIECE WAS RETURNED WITH THE MOUNT AND IMPLANT. FUNCTIONAL TESTING TO RECREATE THE REPORTED EVENT COULD NOT BE PERFORMED DUE TO THE NATURE OF THE DEVICE & EVENT (FRACTURED MOUNT). A PRE-EXISTING CONDITION WAS NOT NOTED ON THE PER FORM. BONE DENSITY WAS UNKNOWN. THE REPORTED DEVICE WAS LOCATED AT AN UNKNOWN TOOTH LOCATION. THE DEVICE WAS IMPLANTED AND EXPLANTED ON THE SAME DAY. X-RAY & PICTURE EVALUATION: THE CUSTOMER DID NOT PROVIDE ANY IMAGES FOR THE REPORTED EVENT. REVIEW OF APPROPRIATE DOCUMENTATION: DOCUMENTS REVIEWED: INSTRUCTIONS FOR USE FOR TAPERED SCREW-VENT®, ADVENT® AND TRABECULAR METAL¿ IMPLANTS 4869 REV 9-10/19. INFORMATION IDENTIFIED: 'CONTRAINDICATIONS' 'WARNINGS' 'CHANGES IN PERFORMANCE' 'ADVERSE EFFECTS.' DHR REVIEW: DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1254577. IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW: COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1254577 FOR SIMILAR EVENT AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: 'DENTAL : FUNCTIONAL : FRACTURE : MOUNT.' POST MARKET TREND REVIEW: JANUARY POST MARKET TRENDING WAS REVIEWED AND THERE WERE NO ACTIONABLE EVENTS OR CORRECTIVE ACTIONS FOR THE REPORTED EVENT OR PRODUCT. NO ACTIONABLE ITEMS HAVE BEEN TRIGGERED THAT WILL AFFECT COMPLAINT HANDLING ON OUR END FOR THIS MONTH. BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION DID OCCUR AND THE REPORTED EVENT WAS CONFIRMED.
ZIMVIE COMPLAINT NUMBER (B)(4).
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
DOCTOR REPORTED MOUNT FRACTURE. NO PATIENT IMPACT. THE PROCEDURE HAS BEEN COMPLETED WITH ANOTHER IMPLANT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 823271 | IMP,TSV,4.1MM,SBM,11.5 | DENTAL IMPLANT | DZE | ZIMMER DENTAL | TSV4B11 | 1254577 | 00889024019232 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Male |