FDA Adverse Event Injury Summary report: N

PITT EASY ABUTMENT

MDR report key: 1624360 · Received March 9, 2010

Report

Report Number
9611820-2010-00002
Event Type
Injury
Date Received
March 9, 2010
Report Date
February 5, 2010
Manufacturer
SYBRON IMPLANT SOLUTIONS GMBH
Product Code
NHA
PMA / PMN Number
K053242
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

IT WAS REPORTED THAT A PATIENT HAD A FRACTURED ABUTMENT HEX. THE PRODUCT WAS RETURNED TO SYBRON IMPLANT SOLUTIONS (B) (4) FOR EVALUATION. THE EVALUATION INDICATED THAT THE ABUTMENT HEX MEETS PRODUCT SPECIFICATIONS AND THAT THE FRACTURE WAS CAUSED BY LOOSENING DUE TO OVERLOADING. THIS IS NOT A PRODUCT FAILURE.

Description of Event or Problem · 1

ON (B) (6) 2010 A DOCTOR REPORTED TO SYBRON IMPLANT SOLUTIONS (B) (4) THAT A PATIENT HAD A PITT EASY ABUTMENT HEX FRACTURE. THIS IS THE FIRST OF TWO INCIDENTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PITT EASY ABUTMENT ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS NHA SYBRON IMPLANT SOLUTIONS GMBH

Patients

Seq Age Sex Outcome Treatment
1 Other| R