FDA Adverse Event Malfunction Summary report: N

MONARCH PLATFORM

MDR report key: 16236540 · Received January 24, 2023

Report

Report Number
3014447948-2023-00003
Event Type
Malfunction
Date Received
January 24, 2023
Date of Event
December 29, 2022
Report Date
February 27, 2023
Manufacturer
AURIS HEALTH, INC.
Product Code
EOQ
UDI-DI
B634MON0000060
PMA / PMN Number
K211493
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

FAILURE ANALYSIS ENGINEERING RECEIVED MULTIPLE IMPACTED PRODUCTS UNDER THE COMPLAINT, AS FOLLOWS: 1.) TWO SCOPES (MBR-000211-A): INVESTIGATIONS WERE COVERED UNDER (B)(4) AND (B)(4), BOTH CONCLUDED THAT THE SCOPES MET SPECIFICATIONS AS REPORTED ISSUE WAS NOT REPLICATED AND NO INCONSISTENT BEHAVIOR OBSERVED. 2.) ONE UMBILICAL CABLE PN (300-0024218-00), INVESTIGATION COVERED UNDER (B)(4). UPON VISUAL INSPECTION, SHOWS DAMAGED/BROKEN AND BENT PINS INSIDE THE LEMO CONNECTOR. FUNCTIONALLY, THE CABLE WAS INSTALLED ON A KNOWN-GOOD-SYSTEM TO TRY TO REPLICATE THE REPORTED ISSUE. SYSTEM BOOT UP SEQUENCE AND SYSTEM INITIALIZATION OPERATED NORMALLY AND AS INTENDED. UPON RUNNING A TEST CASE USING A LUNG MODEL, NAVIGATION REGISTRATION WAS COMPLETED SUCCESSFULLY WITH NO FAULTS OF ANY KINDS. A DRIVE TEST WAS PERFORMED FOR 20 MINUTES, AND DURING THAT PERIOD, NO FAULTS OR ANY INCONSISTENT FUNCTIONAL SYSTEM BEHAVIOR OBSERVED. DURING BENCH TEST, A CONTINUITY TEST WAS PERFORMED BETWEEN THE SUBJECT UMBILICAL CABLE AND THE RETURNED MATING BULKHEAD ASSEMBLY. CONTINUITY TEST FAILED ON BOTH PINS 7 AND 8 (SAME PINS FOUND RECESSED BY VISUAL INSPECTION). THE ASSEMBLY WIRING DIAGRAM FOR THIS CABLE ASSEMBLY WAS REVIEWED, IT WAS CONFIRMED THAT PINS 7 AND 8 ARE PART OF THE ETHERNET COMMUNICATION LINE ¿X2¿ WHICH CONNECTS TO THE TOWER COMMUNICATION SWITCH. ALTHOUGH THE DISCONNECTION WAS CONFIRMED ON THE CABLE, IT MOST LIKELY WOULD NOT CAUSE THE 1200 FAULTS, HOWEVER, THERE WOULD BE A POSSIBILITY THAT A COMMUNICATION/INITIALIZATION FAULT MIGHT BE TRIGGERED. REVIEWING THE LOGS IN S3 (STUDYDATA/2022-12-29/SYSTEM(B)6)_MAN/{6616B7B3-EC43-4D07-9FE9-22DBD6C18D77}/UPLOAD/PROTOLOGS/), IT WAS CONFIRMED THAT THE SYSTEM EXPERIENCED AN ETHERCAT INITIALIZATION FAULT 1405-50-0-0. THE LOGS SHOWED THAT ONLY ONE SLAVE (ADDRESS 1001) WAS DETECTED DURING THE SYSTEM BOOTUP SEQUENCE FOR THE DAY, AFTER SYSTEM REBOOT, SYSTEM INITIALIZED PROPERLY. UPON DIS-ASSEMBLY, IT WAS ALSO NOTICED THAT THERE ARE SOME SCUFF MARKS PRESENT OF THE LEMO CONNECTOR NUT. THIS WAS DISCUSSED WITH THE CABLE¿S SUPPLIER AND CONFIRMED IT IS CAUSED BY THE LEMO FLAT SPANNER BEING WORN-OUT OR HAVE PLAY NATURE WHEN USED. IN-ADDITION, THE PIN TIP FOUND BENT WHICH SUGGEST IT WAS DEFLECTED BEFORE IT GOT RECESSED. THE PINS GETTING RECESSED ON THE UMBILICAL CABLE IS A KNOWN ISSUE, THAT IS DUE TO THE DESIGN OF THE COMPONENT. WHEN THE PINS ARE DEFLECTED AND NOT STRAIGHT, IT WOULD HIT THE MATING PART INSULATION WHICH APPLIES HIGH RETENTION FORCES THE UMBILICAL CABLE PINS ARE UNABLE TO WITHSTAND, LEADING TO PINS BEING RECESSED. BULKHEAD ASSEMBLY, CABLE, CART, UMBILICAL TO COMPUTER AND PDU, SOLDER (300-021048-00). UPON VISUAL INSPECTION, SHOWS DAMAGES TO THE INSULATION DUE TO THE BENT/BROKEN PINS INSIDE THE MATING PART LEMO CONNECTOR (UMBILICAL CABLE) INVESTIGATED UNDER (B)(4). FUNCTIONALLY, THE BULKHEAD ASSEMBLY WAS FULLY FUNCTIONAL DESPITE THE DAMAGES TO THE INSULATION. A MATING TEST WAS PERFORMED USING GOOD-KNOWN- UMBILICAL CABLE TO RE-PRODUCE THE FAILURE. THE BULKHEAD WAS SUBJECTED TO X40 OF MATING/DE-MATING CYCLES. NO DAMAGE FURTHER DAMAGE WAS CREATED ON THE BULKHEAD, NOR ANY DAMAGES OBSERVED OF THE TEST UMBILICAL CABLE. THIS CONFIRMS THAT THE ISSUE CAUSED BY THE ORIGINAL UMBILICAL CABLE THAT WAS RETURNED AND INVESTIGATED ON (B)(4). CONCLUSION: THE REPORTED ISSUE WAS CONFIRMED THROUGH THE LOGS, HOWEVER, WAS NEVER REPLICATED DURING THE INVESTIGATION ON ALL RETURNED PRODUCTS. SCOPES FUNCTIONED AS DESIGNED WITH NO INCONSISTENT BEHAVIOR OBSERVED. THE FAILURE ON THE UMBILICAL CABLE AND BULKHEAD ASSEMBLY IS A KNOWN ISSUE, THAT IS DUE TO THE ASSEMBLY PROCESS OF THE COMPONENT (LEMO CONNECTOR), WHEN THE PINS ARE DEFLECTED AND NOT STRAIGHT, IT WOULD HIT THE MATING PART INSULATION WHICH APPLIES HIGH RETENTION FORCES THE UMBILICAL CABLE PINS ARE UNABLE TO WITHSTAND, LEADING TO PINS BEING RECESSED. A REVIEW OF THE SYSTEM HISTORY RECORD (DHR) SHOWED NO NON-CONFORMANCES REPORTED THAT IS RELATED TO THE REPORTED EVENT. THIS ISSUE WILL BE TRACKED AND TRENDED IN THE MONTHLY COMPLAINT TREND REVIEW MEETINGS.

Description of Event or Problem · 0

DURING A MONARCH BRONCHOSCOPY PROCEDURE IT WAS REPORTED THE CUSTOMER HAD TWO BRONCHOSCOPES IN A ROW THAT PRODUCED FAULT ID: (B)(4) , POOR QUALITY SCOPE EM DATA DETECTED DURING NAVIGATION (NAV) REGISTRATION. THE FIRST INSTANCE OCCURRED DURING NAV REGISTRATION IMMEDIATELY AFTER PRESSING SET ROLL CORRECTION. THE CUSTOMER PROACTIVELY POWER-CYCLED THE SYSTEM AND SWITCHED TO A SECOND BRONCHOSCOPE; HOWEVER, THE ERROR PERSISTED. THE CUSTOMER COULDN¿T PERFORM THE MONARCH PROCEDURE AND THE CASE WAS ABORTED. THERE WAS NO REPORTED HARM TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
638493 MONARCH PLATFORM BRONCHOSCOPE (FLEXIBLE OR RIGID) AND ACCESSORIES EOQ AURIS HEALTH, INC. MON-000006 B634MON0000060

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other