FDA Adverse Event Malfunction Summary report: N

G.E. MEDICAL SYSTEMS

MDR report key: 1623579 · Received March 3, 2010

Report

Report Number
MW5014993
Event Type
Malfunction
Date Received
March 3, 2010
Date of Event
February 23, 2010
Report Date
February 26, 2010
Product Code
MHX
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
WI, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

G.E. DASH 5000 PHYSIOLOGIC MONITOR LOST ECG TRACING ON STABLE POST-OPERATIVE PATIENT WHO HAD BEEN IN A NORMAL SINUS RHYTHM. "LEAD FAIL" MESSAGE REGISTERED ON SCREEN. ALL LEADS WERE ATTACHED TO PATIENT SECURELY. APPARENT HISTORIC 2-3 YEARS LONG PROBLEM OF "LEADS FAIL" MESSAGE WITH CORRESPONDENT LOSS OF ECG TRACING HAS BEEN DEALT WITH MANUFACTURER'S REP WITHOUT SUCCESSFUL RESOLUTION TO DATE. ONE OF THESE 10 MONITORS WAS EVEN TAKEN TO BE BENCH-TESTED AT G.E. MEDICAL; WHERE IT WAS DETERMINED TO PERFORM NOMINALLY. THIS APPEARS TO BE AN INTERMITTENT PROBLEM THAT 13 UNPREDICTABLE. THE NURSES AND TECHS USING THIS EQUIPMENT HAVE ALL BEEN PROPERLY IN-SERVICED IN USE OF THESE MONITORS. THIS HOSPITAL IS FILLED WITH SIMILAR (ALTHOUGH NOT SAME) MONITORS - AND WE HAVE NOT HEARD OF SIMILAR REPORTS OF PROBLEMS AS WE HAVE WITH THESE DASH 5000 SERIES MONITORS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 G.E. MEDICAL SYSTEMS DASH 5000 PHYSIOLOGIC MONITOR MHX DASH 5000
10 G.E. MEDICAL SYSTEMS DASH 5000 PHYSIOLOGIC MONITOR MHX DASH 5000
2 G.E. MEDICAL SYSTEMS DASH 5000 PHYSIOLOGIC MONITOR MHX DASH 5000
3 G.E. MEDICAL SYSTEMS DASH 5000 PHYSIOLOGIC MONITOR MHX DASH 5000
4 G.E. MEDICAL SYSTEMS DASH 5000 PHYSIOLOGIC MONITOR MHX DASH 5000
5 G.E. MEDICAL SYSTEMS DASH 5000 PHYSIOLOGIC MONITOR MHX DASH 5000
6 G.E. MEDICAL SYSTEMS DASH 5000 PHYSIOLOGIC MONITOR MHX DASH 5000
7 G.E. MEDICAL SYSTEMS DASH 5000 PHYSIOLOGIC MONITOR MHX DASH 5000
8 G.E. MEDICAL SYSTEMS DASH 5000 PHYSIOLOGIC MONITOR MHX DASH 5000
9 G.E. MEDICAL SYSTEMS DASH 5000 PHYSIOLOGIC MONITOR MHX DASH 5000

Patients

Seq Age Sex Outcome Treatment
1