FDA Adverse Event
Malfunction
Summary report: N
CARDIOSAVE HYBRID TYPE B PLUG
MDR report key: 16234010
·
Received January 24, 2023
Report
- Report Number
- 2249723-2023-00397
- Event Type
- Malfunction
- Date Received
- January 24, 2023
- Date of Event
- September 11, 2020
- Report Date
- January 17, 2023
- Manufacturer
- DATASCOPE CORP. - MAHWAH
- Product Code
- DSP
- UDI-DI
- 10607567108391
- PMA / PMN Number
- K112372
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO
- Reporter Occupation
- 003
Narratives
Additional Manufacturer Narrative · 0
THE STM REPLACED THE HELIUM ACCESS PANEL (0333-00-0245-02) THEN COMPLETED PM SERVICE. ALL SAFETY, FUNCTIONALITY, AND CALIBRATION CHECKS WERE PERFORMED AND ALL TESTS PASSED TO FACTORY SPECIFICATIONS. THE IABP UNIT WAS CLEARED FOR CLINICAL USE AND RELEASED TO THE CUSTOMER. THIS REPORT IS BEING SUBMITTED AS THE RESULT OF A RETROSPECTIVE REVIEW CONDUCTED IN CAPA (B)(4).
Description of Event or Problem · 0
IT WAS REPORTED THAT DURING PREVENTATIVE MAINTENANCE (PM) PERFORMED BY A GETINGE SERVICE TERRITORY MANAGER (STM), THE HELIUM ACCESS PANEL FOR THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) WAS OBSERVED TO BE DAMAGED/CRACKED. IT WAS NOTED THAT THE IABP UNIT INVOLVED IS A RENTAL UNIT. THERE WAS NO PATIENT INVOLVEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1403320 | CARDIOSAVE HYBRID TYPE B PLUG | SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL | DSP | DATASCOPE CORP. - MAHWAH | 0998-00-0800-53 | 10607567108391 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |