FDA Adverse Event Malfunction Summary report: N

FREESTYLE FREEDOM LITE

MDR report key: 1623350 · Received March 5, 2010

Report

Report Number
2954323-2010-00258
Event Type
Malfunction
Date Received
March 5, 2010
Date of Event
February 11, 2010
Report Date
March 5, 2010
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED METER ((B) (4)) WAS RETURNED, BUT NOT THE TEST STRIPS FROM LOT # 0918834. TESTS STRIPS FROM LOT # 0915618 WERE RETURNED INSTEAD. THE COMPLAINT WAS NOT CONFIRMED. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION DURING CONTROL SOLUTION TESTING. THE REPORTED READINGS WERE FOUND IN THE METER MEMORY LOG WITHIN A TEN MINUTE TIMEFRAME (DATE: (B) (6) 2010). THIS IS THE FINAL REPORT.

Description of Event or Problem · 1

A CUSTOMER REPORTED SHE RECEIVED THE FOLLOWING READINGS ON HER BLOOD GLUCOSE METER WITHIN 10 MINUTES: 41 MG/DL, 170 MG/DL, 439 MG/DL, 160 MG/DL AND 173 MG/DL. RESULTS WERE PLOTTED ON A PARKES ERROR GRID AND FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FREESTYLE FREEDOM LITE BLOOD GLUCOSE MONITORING SYSTEM NBW 0918834

Patients

Seq Age Sex Outcome Treatment
1