AVANTA SINGLE-PATIENT DISPOSABLE SET
Report
- Report Number
- 2520313-2010-00012
- Event Type
- Death
- Date Received
- March 3, 2010
- Date of Event
- February 1, 2010
- Report Date
- March 3, 2010
- Manufacturer
- MEDRAD, INC.
- Product Code
- DXT
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHARMACIST
Narratives
A MEDRAD SERVICE ENGINEER PERFORMED A SYSTEM CHECK-OUT ON THE AVANTA FLUID MANAGEMENT INJECTION SYSTEM AND THE UNIT WAS FOUND TO BE OPERATING TO SPECIFICATION. THE SINGLE-PATIENT DISPOSABLE SET (SPAT) THAT WAS IN-USE DURING THE REPORTED EVENT WAS RETAINED BY THE CUSTOMER. WE ARE AWAITING THE RETURN OF THE DISPOSABLE FOR EVALUATION. A FOLLOW-UP REPORT WILL BE SUBMITTED, ONCE THE DISPOSABLE IS RETURNED AND WE EVALUATE IT. ADDITIONAL APPLICATIONS TRAINING WAS PERFORMED OVER MULTIPLE DAYS AT THE SITE. DURING THE TRAINING, WE OBSERVED THAT THE STAFF WAS NOT PROPERLY PURGING THE DISPOSABLES AND FOLLOWING THE INSTRUCTIONS THAT WERE PROVIDED DURING TRAINING (WHICH ARE ALSO CONTAINED IN THE OPERATOR'S MANUAL). WE ATTEMPTED TO ADDRESS AND CORRECT THIS DURING THE ADDITIONAL APPLICATIONS TRAINING. IN ADDITION, DISPOSABLES FROM THE SAME LOT NUMBER THAT WERE IN-USE DURING THE REPORTED EVENT WERE USED DURING THE ADDITIONAL APPLICATIONS TRAINING WITHOUT INCIDENT. AFTER THE AVANTA INJECTOR WAS CHECKED BY MEDRAD SERVICE AND AFTER THIS APPLICATIONS TRAINING, THE SITE CONTINUED TO USE THE INJECTOR WITHOUT INCIDENT.
THE SITE REPORTED THE FOLLOWING: DURING THE FIRST INJECTION OF THE PROCEDURE ON THE THIRD PATIENT OF THE DAY, AIR WAS INJECTED INTO THE LEFT CORONARY ARTERY. THE PATIENT WENT INTO HEART AND RESPIRATORY FAILURE, AND REQUIRED REANIMATION AND CIRCULATORY ASSISTANCE. THE PATIENT WAS THEN TRANSFERRED TO THE INTENSIVE CARE UNIT AND EXPIRED THE NEXT DAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AVANTA SINGLE-PATIENT DISPOSABLE SET | ANGIOGRAPHIC TUBING SET | DXT | MEDRAD, INC. | 3018355 | 720005 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR | Death | AVANTA FLUID MANAGEMENT INJECTION SYSTEM |