FDA Adverse Event Malfunction Summary report: N

EON MINI IPG

MDR report key: 1623086 · Received February 24, 2010

Report

Report Number
1627487-2009-00096
Event Type
Malfunction
Date Received
February 24, 2010
Date of Event
August 25, 2009
Report Date
August 25, 2009
Manufacturer
ADVANCED NEUROMODULATION SYSTEMS, INC.
Product Code
LGW
PMA / PMN Number
P010032
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

"MALFUNCTION" IS INTERPRETED AS A COMPROMISE OF THE DEVICE'S THERAPEUTIC EFFECTIVENESS (LOSS OF PAIN RELIEF) WHICH CONTRIBUTED TO SIGNIFICANT DEVICE ADVERSE EVENT RESULTING IN A SURGICAL PROCEDURE TO REMOVE THE DEVICE. EVALUATION: DEVICE HISTORY AND STERILIZATION RECORDS WERE REVIEWED. RESULTS: THE DEVICE HISTORY AND STERILIZATION RECORDS REVIEWED WERE FOUND TO MEET SPECIFICATIONS AND NO ANOMALIES WERE FOUND. CONCLUSION: THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION RECORDS. THIS MDR IS BEING SUBMITTED PAST THE 30 DAY REPORTING REQUIREMENT AS PART OF A RETROSPECTIVE REVIEW INITIATED IN RESPONSE TO AN FDA INSPECTION. A RETROSPECTIVE REVIEW OF THE COMPLAINT RECORD DETERMINED THAT ANS MISINTERPRETED THE MDR REGULATIONS IN THIS INSTANCE. ANS HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. ANS DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

THE PT RECEIVED HIS IPG ON (B) (6) 2009. IT WAS REPORTED THAT ABOUT A MONTH POSTOPERATIVELY, THE IPG WOULD NOT HOLD A CHARGE FOR MORE THAN ABOUT 24 HOURS. PT ALSO WAS REPORTED TO HAVE DEVELOPED (B) (6) INFECTION POST-OPERATIVELY AND WAS TREATED WITH ANTIBIOTICS. THE PHYSICIAN EXPLANTED BUT DID NOT REPLACE THE IPG ON (B) (6) 2009.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EON MINI IPG TOTALLY IMPLANTABLE PULSE GENERATOR LGW ADVANCED NEUROMODULATION SYSTEMS, INC. 3788 2796044

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention