EON MINI IPG
Report
- Report Number
- 1627487-2009-00335
- Event Type
- Malfunction
- Date Received
- February 24, 2010
- Date of Event
- September 24, 2009
- Report Date
- October 23, 2009
- Manufacturer
- ADVANCED NEUROMODULATION SYSTEMS, INC.
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
"MALFUNCTION" IS INTERPRETED AS A COMPROMISE OF THE DEVICE'S THERAPEUTIC EFFECTIVENESS (LOSS OF PAIN RELIEF) WHICH CONTRIBUTED TO SIGNIFICANT DEVICE ADVERSE EVENT RESULTING IN A SURGICAL PROCEDURE TO REMOVE THE DEVICE. EVALUATION RESULTS: THE DEVICE HISTORY AND STERILIZATION RECORDS REVIEWED WERE FOUND TO MEET SPECIFICATIONS AND NO ANOMALIES WERE FOUND. CONCLUSION: THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION RECORDS. THIS MDR IS BEING SUBMITTED PAST THE 30 DAY REPORTING REQUIREMENT AS PART OF A RETROSPECTIVE REVIEW INITIATED IN RESPONSE TO AN FDA INSPECTION. A RETROSPECTIVE REVIEW OF THE COMPLAINT RECORD DETERMINED THAT ANS MISINTERPRETED THE MDR REGULATIONS IN THIS INSTANCE. ANS HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. ANS DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.
THE PT RECEIVED HER IPG ON (B) (6) 2009. IT WAS REPORTED THAT ON (B) (6) 2009, THE PT SAID HER IPG BATTERY APPEARED TO BE DEPLETING MORE QUICKLY THAN IT HAD BEEN. PER HER PARAMETERS, IT WAS ESTIMATED THAT THE IPG SHOULD LAST ABOUT 20 DAYS BETWEEN RECHARGING. THE PT WAS NOW RECHARGING ABOUT EVERY 4 DAYS. THE PHYSICIAN EXPLANTED AND REPLACED THE IPG ON (B) (6) 2009. FOLLOW-UP ON THE PT FOUND THAT SHE IS DOING WELL WITH NO FURTHER ISSUES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | EON MINI IPG | TOTALLY IMPLANTABLE PULSE GENERATOR | LGW | ADVANCED NEUROMODULATION SYSTEMS, INC. | 3788 | 2770076 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |