Report
- Report Number
- 1627487-2009-00092
- Event Type
- Malfunction
- Date Received
- February 24, 2010
- Date of Event
- April 17, 2009
- Report Date
- April 17, 2009
- Manufacturer
- ADVANCED NEUROMODULATION SYSTEMS, INC.
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
"MALFUNCTION" IS INTERPRETED AS A COMPROMISE OF THE DEVICE'S THERAPEUTIC EFFECTIVENESS (LOSS OF PAIN RELIEF) WHICH CONTRIBUTED TO SIGNIFICANT DEVICE ADVERSE EVENT RESULTING IN A SURGICAL PROCEDURE TO REMOVE THE DEVICE. EVAL RESULTS: THE DEVICE HISTORY AND STERILIZATION RECORDS REVIEWED WERE FOUND TO MEET SPECIFICATIONS AND NO ANOMALIES WERE FOUND. CONCLUSION: THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION RECORDS. THIS MDR IS BEING SUBMITTED PAST THE 30 DAY REPORTING REQUIREMENT AS PART OF A RETROSPECTIVE REVIEW INITIATED IN RESPONSE TO AN FDA INSPECTION. A RETROSPECTIVE REVIEW OF THE COMPLAINT RECORD DETERMINED THAT ANS MISINTERPRETED THE MDR REGULATIONS IN THIS INSTANCE. ANS HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. ANS DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.
THE PT RECEIVED HER IPG ON (B) (6) 2008. IT WAS REPORTED THAT THE IPG WOULD ONLY HOLD A CHARGE 8 TO 10 HOURS. PREVIOUSLY THE CHARGE WOULD LAST FOR 3 TO 4 DAYS. THE PT DOES HAVE HIGH PARAMETERS, BUT HAS NOT BEEN REPROGRAMMED. PT TRIED 2 DIFFERENT CHARGING SYSTEMS WITH THE SAME RESULTS. THE PHYSICIAN EXPLANTED THE IPG AND REPLACED IT WITH THE SAME MODEL IPG. FOLLOW UP WITH THE PT FOUND THAT SHE IS DOING VERY WELL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | EON MINI IPG | IMPLANTABLE PULSE GENERATOR (IPG) | LGW | ADVANCED NEUROMODULATION SYSTEMS, INC. | 3788 | 172558 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |