BD PHOENIX¿ PMIC/ID-107
Report
- Report Number
- 1119779-2023-00056
- Event Type
- Malfunction
- Date Received
- January 23, 2023
- Date of Event
- November 28, 2022
- Report Date
- May 25, 2023
- Manufacturer
- BECTON, DICKINSON & CO. (SPARKS)
- Product Code
- LON
- UDI-DI
- 30382904486076
- PMA / PMN Number
- K040106
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H6. INITIAL SUMMARY: THIS COMPLAINT IS NOT CONFIRMED. THIS COMPLAINT IS FOR MISIDENTIFICATION WHEN USING PHOENIX PANEL PMIC/ID-107 (448607) BATCH NUMBERS 2060227, 2207540 AND 0130382. BATCH 0130382 DOES NOT EXIST FOR THIS PANEL MATERIAL #448607 AND WAS NOT INVESTIGATED. THE CUSTOMER DID NOT RETURN PANELS OR PHOENIX GENERATED LAB REPORTS BUT PROVIDED FOUR (4) ISOLATES FOR THE INVESTIGATION. THE FOUR (4) CUSTOMER RETURNED ISOLATES WERE LABELED AS FL-1, FL-3, FL-5, AND FL-6. TO INVESTIGATE, EACH OF THE FOUR (4) CUSTOMER RETURNED ISOLATES WERE TESTED ON A PHOENIX M50 WITH A RETENTION PANELS FROM EACH THE COMPLAINT BATCHES 2060227 AND 2207540 FOR A TOTAL OF (8) PANELS TESTED AND EVALUATED FOR PHOENIX IDENTIFICATION RESULTS. FOR FURTHER INVESTIGATION, EACH CUSTOMER RETURNED ISOLATE FL-1, FL-3, FL-5, FL-6 WERE RAN ON A BRUKER MALDI FOR IDENTIFICATION RESULTS. BRUKER MALDI PROVIDED THE FOLLOWING RESULTS: ISOLATE FL-1 WAS IDENTIFIED AS S. DYSGALACTIAE ON THE BRUKER MALDI. IT IS TO BE NOTED THAT S. DYSGALACTIAE IS A MEMBER OF THE LANCEFIELD GROUPS C AND G. ISOLATE FL-3 WAS IDENTIFIED AS S. EPIDERMIDIS. ISOLATES FL-5 AND FL-6 WERE IDENTIFIED AS S. SAPROPHYTICUS. DURING THE INVESTIGATION, ALL EIGHT (8) PANELS TESTED YIELDED EXPECTED IDENTIFICATION RESULTS. THEREFORE, THIS COMPLAINT IS NOT CONFIRMED. A REVIEW OF QUALITY NOTIFICATIONS REVEALED NO QUALITY NOTIFICATIONS GENERATED ON EITHER OF THE COMPLAINT BATCHES 2207540 AND 2060227. A REVIEW OF COMPLAINTS REVEALED THREE (3) ADDITIONAL COMPLAINTS ON BATCH 2207540. NONE OF THE 3 ADDITIONAL COMPLAINTS ARE RELATED TO THIS DEFECT AND NONE HAVE BEEN CONFIRMED. A REVIEW OF COMPLAINTS REVEALED SEVEN (7) ADDITIONAL COMPLAINTS ON BATCH 2060227. OF THE SEVEN (7) ADDITIONAL COMPLAINTS, ONE (1) IS RELATED TO THIS DEFECT (MIS ID OF S. SAPROPHYTICUS) AND HAS BEEN CONFIRMED. A REVIEW OF QUALITY NOTIFICATIONS COULD NOT BE PERFORMED FOR BATCH 0130382 SINCE THE BATCH NUMBER WAS INVALID. A REVIEW OF COMPLAINTS FOR BATCH 0130382 COULD NOT BE PERFORMED SINCE THE BATCH NUMBER WAS INVALID. COMPLAINT TRENDING WAS PERFORMED AND NO TRENDS WERE IDENTIFIED ASSOCIATED WITH THIS DEFECT (MIS ID OF S. SAPROPHYTICUS) . BD ID/AST PLANT QUALITY WILL CONTINUE TO MONITOR FOR TRENDS AND TAKE ACTION AS NECESSARY.
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
REPORTED 4 OF 5. IT WAS REPORTED THAT BD PHOENIX¿ PMIC/ID-107 MISIDENTIFICATION OCCURRED USING SEVERAL CLINICAL SAMPLES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER IS REPORTING MISIDENTIFICATION OF ORGANISMS USING 448607, LOT NUMBER 0130382 (B)(6) 22-PHOENIX M50 ID AS CORYNE JEIKEIUM - REPEATED ON PHOENIX M50 AS STA EPI.
REPORTED 4 OF 5 IT WAS REPORTED THAT BD PHOENIX¿ PMIC/ID-107 MISIDENTIFICATION OCCURRED USING SEVERAL CLINICAL SAMPLES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER IS REPORTING MISIDENTIFICATION OF ORGANISMS USING 448607, LOT NUMBER 0130382 (B)(6)2022 -PHOENIX M50 ID AS (B)(6) - REPEATED ON PHOENIX M50 AS STA EPI
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 484003 | BD PHOENIX¿ PMIC/ID-107 | SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY | LON | BECTON, DICKINSON & CO. (SPARKS) | 448607 | 0130382 | 30382904486076 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |