FDA Adverse Event Other Summary report: N

ADVIA CENTAUR HBC TOTAL (HBCT)

MDR report key: 1622439 · Received February 19, 2010

Report

Report Number
1219913-2010-00020
Event Type
Other
Date Received
February 19, 2010
Date of Event
February 4, 2010
Report Date
February 5, 2010
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC.
Product Code
LOM
PMA / PMN Number
P040004
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE CAUSE FOR THE FALSE NEGATIVE (B)(6) TOTAL RESULT IS UNKNOWN. A SIEMENS REPRESENTATIVE EXAMINED THE (B)(6) TOTAL QC AND CALIBRATIONS. NO ISSUES WERE FOUND. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Additional Manufacturer Narrative · 1

SIEMENS FILED THE INITIAL MDR ON FEBRUARY 19, 2010. (B)(6) 2012: ADDITIONAL INFORMATION: (B)(6) 2010: THE SAMPLE IS OF A PATIENT WITH A KNOWN PAST INFECTION AND NOT A POSITIVE POOL INTERNAL CONTROL SAMPLE AS WAS STATED IN THE INITIAL MDR. THE CUSTOMER USES THIS SAMPLE AS AN ADDITIONAL INTERNAL CONTROL. (B)(6) FOR THE SAMPLE. (B)(6) 2010: THE CUSTOMER SENT AN ALIQUOT OF THE PATIENT SAMPLE TO SIEMENS HEALTHCARE DIAGNOSTICS FOR FURTHER TESTING. THE COLLECTIVE DATA GENERATED AT THE CUSTOMER SITE AND AT SIEMENS SUGGESTS THAT THE PATIENT SAMPLE USED AS A SECONDARY CONTROL BY THE CUSTOMER WILL MEASURE BOTH ABOVE AND BELOW THE ASSAY CUT OFF. THE ROOT CAUSE IS UNKNOWN.

Description of Event or Problem · 1

A FALSE NEGATIVE ADVIA CENTAUR (B)(6) TOTAL RESULT WAS OBTAINED ON A POSITIVE POOL INTERNAL CONTROL SAMPLE. THE CUSTOMER ALWAYS RUNS A POSITIVE POOL SAMPLE FOR (B)(6). THE POSITIVE POOL SAMPLE WAS TESTED WITH A DIFFERENT METHOD AND THE RESULT WAS POSITIVE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADVIA CENTAUR HBC TOTAL (HBCT) HBCT IMMUNOASSAY LOM SIEMENS HEALTHCARE DIAGNOSTICS INC. NA 032

Patients

Seq Age Sex Outcome Treatment
1