FDA Adverse Event Malfunction Summary report: N

CS100 INTRA-AORTIC BALLOON PUMP, CHINESE, 220V

MDR report key: 16223119 · Received January 23, 2023

Report

Report Number
2249723-2023-00278
Event Type
Malfunction
Date Received
January 23, 2023
Date of Event
January 8, 2023
Report Date
January 30, 2025
Manufacturer
DATASCOPE CORP. - MAHWAH
Product Code
DSP
UDI-DI
10607567107790
PMA / PMN Number
K031636
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF OUR INVESTIGATION. DUE TO CHARACTER LIMITATION COMPLETE E1(SITE NAME) - HUAZHONG UNIVERSITY OF SCIENCE TONGJI HOSPITAL, TONGJI MEDICAL COLLEGE.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: E1(VENT SITE POSTAL CODE: (B)(6)). A GETINGE FIELD SERVICE ENGINEER CONFIRMED SINCE THE CUSTOMER'S REPAIR PLAN AND PRICE HAVE NOT YET BEEN CONFIRMED, HIS REPAIR WORK HAS NOT BEEN CARRIED OUT. THE EQUIPMENT IS CURRENTLY OUT OF SERVICE AND AWAITING REPAIR."

Additional Manufacturer Narrative · 0

A GETINGE FIELD SERVICE ENGINEER (FSE) WHEN THE MACHINE IS STARTED, THE EQUIPMENT REPORTS A LOOP LEAKAGE, ENTERS THE MAINTENANCE MODE FOR TESTING, AND LEAKS ARE FOUND. THE PRESSURE CANNOT BE MAINTAINED, AND IT IS JUDGED THAT MAINTENANCE IS NEEDED. AFTER REPLACING THE 5000-HOUR MAINTENANCE KIT, ALL PARAMETERS ARE RETESTED AND NORMAL, CONFIRMING THAT THE FAULT HAS BEEN REPAIRED AND CAN BE USED CLINICALLY.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING ROUTINE CHECK, THE CS100 INTRA-AORTIC BALLOON PUMP (IABP) JAD A LOOP AIR LEAKAGE ALARM. THERE WAS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 0

N/A.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
432776 CS100 INTRA-AORTIC BALLOON PUMP, CHINESE, 220V SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL DSP DATASCOPE CORP. - MAHWAH 0998-00-3013-45 10607567107790

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown