BD¿ AMBER ORAL SYRINGE WITH NON LUER TIP, 3 ML
Report
- Report Number
- 1213809-2023-00022
- Event Type
- Malfunction
- Date Received
- January 20, 2023
- Date of Event
- January 3, 2023
- Report Date
- February 23, 2023
- Manufacturer
- BECTON DICKINSON MEDICAL SYSTEMS
- Product Code
- KYW
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. DEVICE MANUFACTURE DATE: UNKNOWN. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
H6: INVESTIGATION SUMMARY TWO PHOTOS WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. BASED ON THE INVESTIGATION AND WITH THE RETURNED PHOTO SAMPLE ANALYSIS THE SYMPTOM REPORTED IS CONFIRMED. POTENTIAL ROOT CAUSE FOR THE MIXED PRODUCT DEFECT IS ASSOCIATED WITH INADEQUATE LINE CLEARANCE IN THE PACKAGING PROCESS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 1082858. A REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. H3 OTHER TEXT : SEE H10
IT WAS REPORTED WHILE USING BD¿ AMBER ORAL SYRINGE WITH NON LUER TIP, 3 ML THERE WERE MIXED PRODUCTS IN THE PACK. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: A DIFFERENT MODEL OF SYRINGE FOUND WITHIN AMBER ORAL SYRINGE LOT 1082858.
IT WAS REPORTED WHILE USING BD¿ AMBER ORAL SYRINGE WITH NON LUER TIP, 3 ML THERE WERE MIXED PRODUCTS IN THE PACK. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: A DIFFERENT MODEL OF SYRINGE FOUND WITHIN AMBER ORAL SYRINGE LOT 1082858.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2104345 | BD¿ AMBER ORAL SYRINGE WITH NON LUER TIP, 3 ML | LIQUID MEDICATION DISPENSER | KYW | BECTON DICKINSON MEDICAL SYSTEMS | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |