T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY
MDR report key: 16215919
·
Received January 20, 2023
Report
- Report Number
- 3013756811-2023-09296
- Event Type
- Malfunction
- Date Received
- January 20, 2023
- Date of Event
- December 29, 2022
- Report Date
- January 20, 2023
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- QFG
- UDI-DI
- 00850006613380
- PMA / PMN Number
- K201214
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 0
NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE CARTRIDGE USER GUIDE CAUTIONS THAT INSULIN SHOULD BE AT ROOM TEMPERATURE BEFORE FILLING A CARTRIDGE. THE TANDEM USER GUIDE INSTRUCTS THE USER TO REMOVE ANY RESIDUAL AIR FROM THE CARTRIDGE.
Description of Event or Problem · 0
IT WAS REPORTED THAT THE INSULIN GAUGE WAS INACCURATE. REPORTEDLY, THE CUSTOMER LOADED COLD INSULIN INTO THE CARTRIDGE AND WAS NOT PERFORMING THE AIR REMOVAL STEP. CUSTOMER LOADED A NEW CARTRIDGE TO RESOLVE THE ISSUE. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 120-174 MG/DL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2005861 | T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY | ALTERNATE CONTROLLER ENABLED INFUSION PUMP | QFG | TANDEM DIABETES CARE | 1002717 | W0538917 | 00850006613380 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | Male |