FDA Adverse Event Malfunction Summary report: N

BD BBL¿ TAXO¿ DIFFERENTIATION DISCS FOR V FACTOR

MDR report key: 16209274 · Received January 19, 2023

Report

Report Number
1119779-2023-00045
Event Type
Malfunction
Date Received
January 19, 2023
Date of Event
January 10, 2023
Report Date
February 20, 2023
Manufacturer
BECTON, DICKINSON & CO. (SPARKS)
Product Code
JTO
UDI-DI
00382902317272
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H6 INVESTIGATION SUMMARY: A COMPLAINT INVESTIGATION DUE TO NO GROWTH FOR TAXO V FACTORS CATALOG 231727 BATCH NO.: 6924935 WAS PERFORMED ON RETENTION SAMPLES. RETURNED GOODS WERE NOT RECEIVED FROM CUSTOMER. THE INVESTIGATION REQUIRED TO TEST PRODUCT FOR PERFORMANCE, VISUAL INSPECTION, AND BATCH RECORD REVIEW. RETENTION SAMPLES PERFORMED AS EXPECTED. NO DISCREPANCIES OBSERVED DURING THE VISUAL INSPECTION. BATCH RECORD REVIEW DID NOT SHOW ANY EVIDENCE OF CUSTOMER CLAIM. NO CORRECTIVE ACTION IS REQUIRED BASED ON INFORMATION EVALUATED AND SATISFACTORY RESULTS OBTAINED DURING THE QC TEST. COMPLAINT IS NOT CONFIRMED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT 60 BD BBL¿ TAXO¿ DIFFERENTIATION DISCS FOR V FACTOR DURING VALIDATION NO GROWTH OCCURRED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NO GROWTH OF HAEMOPHILUS INFLUENZAE AT THE EDGE OF 60 DISC WITH THE USE OF A COS + FACTOR V MEDIUM DURING VALIDATION NO GROWTH , ARE TESTING NOW WITH OTHER BATCH , TOMORROW NOON 11/1/2023 12H I HAVE TO CALL THE CUSTOMER FOR RESULTS OTHER BATCH.

Description of Event or Problem · 0

IT WAS REPORTED THAT 60 BD BBL¿ TAXO¿ DIFFERENTIATION DISCS FOR V FACTOR DURING VALIDATION NO GROWTH OCCURRED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NO GROWTH OF HAEMOPHILUS INFLUENZAE AT THE EDGE OF 60 DISC WITH THE USE OF A COS + FACTOR V MEDIUM DURING VALIDATION NO GROWTH , ARE TESTING NOW WITH OTHER BATCH , TOMORROW NOON 11/1/23 12H I HAVE TO CALL THE CUSTOMER FOR RESULTS OTHER BATCH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
389713 BD BBL¿ TAXO¿ DIFFERENTIATION DISCS FOR V FACTOR DISCS, STRIPS AND REAGENTS, MICROORGANISM DIFFERENTIATION JTO BECTON, DICKINSON & CO. (SPARKS) 231727 2056543 00382902317272

Patients

Seq Age Sex Outcome Treatment
1 Unknown