FDA Adverse Event Malfunction Summary report: N

AMSCO 400 STERILIZER

MDR report key: 16204459 · Received January 19, 2023

Report

Report Number
3005899764-2023-00002
Event Type
Malfunction
Date Received
January 19, 2023
Date of Event
December 19, 2022
Report Date
January 19, 2023
Manufacturer
STERIS MEXICO, S. DE R.L. DE C.V.
Product Code
FLE
UDI-DI
00724995210991
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A STERIS SERVICE TECHNICIAN INSPECTED THE STERILIZER AND FOUND NO ISSUES WITH THE FUNCTION OR OPERATION OF THE UNIT. THE TECHNICIAN CHECKED ALL CYCLE PARAMETERS, RAN DART AND LEAK TEST CYCLES AND FOUND TEMPERATURE AND WATER PRESSURE TO ALL BE WITHIN SPECIFICATION. THE TECHNICIAN RETURNED THE STERILIZER TO SERVICE AND NO ADDITIONAL ISSUES HAVE BEEN REPORTED.

Additional Manufacturer Narrative · 0

THE CUSTOMER IS CONDUCTING AN INVESTIGATION AS A FOLLOW-UP TO "INFLAMMATION" THAT OCCURRED TO A PATIENT POST PROCEDURE. THE USER FACILITY DID NOT DISCLOSE THE CURRENT PATIENT STATUS OR IF ANY MEDICAL TREATMENT WAS ADMINISTERED. WE UNDERSTAND THE USER FACILITY IS REVIEWING ALL THEIR PRE- AND POST-PROCEDURAL STANDARD OPERATING PROCEDURES. THE CUSTOMER CONTACTED STERIS ASKING WE PERFORM AN EVALUATION OF THEIR EQUIPMENT IN AN ATTEMPT TO RULE OUT THE AMSCO 400 STERILIZER AS A CONTRIBUTOR TO THE REPORTED EVENT. THE USER FACILITY DID NOT DISCLOSE THE SPECIFIC INSTRUMENTS THAT WERE USED DURING THE PROCEDURE SUBJECT OF THE REPORTED EVENT. STERIS IS WORKING TO CONDUCT INSPECTION OF THE AMSCO 400 STERILIZER. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE THE INSPECTION IS COMPLETE.

Description of Event or Problem · 0

THE USER FACILITY CONTACTED STERIS REQUESTING FUNCTIONAL TESTING OF THEIR AMSCO 400 STERILIZER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
362942 AMSCO 400 STERILIZER STERILIZER FLE STERIS MEXICO, S. DE R.L. DE C.V. AMSCO 400 STEAM STERILIZER 00724995210991

Patients

Seq Age Sex Outcome Treatment
1 Unknown