ZOLL IVTM ICY CATHETER
Report
- Report Number
- 3010617000-2023-00052
- Event Type
- Injury
- Date Received
- January 18, 2023
- Date of Event
- September 18, 2022
- Report Date
- January 18, 2023
- Manufacturer
- ZOLL CIRCULATION
- Product Code
- NCX
- PMA / PMN Number
- K101987
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE ICY CATHETER ASSOCIATED WITH THIS COMPLAINT WAS NOT RETURNED FOR EVALUATION. THE CUSTOMER DISPOSED THE CATHETER. SINCE THE DEVICE WAS NOT RETURNED, AN INVESTIGATION COULD NOT BE PERFORMED AND A ROOT CAUSE COULD NOT BE DETERMINED. EVENT OF DVT WAS ASSESSED AS SERIOUS BECAUSE BASED ON AVAILABLE INFORMATION, TREATMENT WAS REQUIRED. EVENT ASSESSED AS PROBABLY RELATED TO THE ZOLL CATHETER DUE TO RELEVANT TIMING AND LOCATION OF DVT. AT THE SAME TIME, THE PATIENT'S POST CARDIAC ARREST CONDITION AND IMMOBILITY POSSIBLY CONTRIBUTED IN DEVELOPMENT OF CURRENT CONDITION OF DVT. DVT IS A KNOWN COMPLICATION OF CENTRAL CATHETERS OF ANY KIND. PATIENTS IN A CRITICAL CONDITION ARE TREATED WITH IVTM. IN SUCH CASES, THOSE CONDITIONS MAKE THE PATIENTS PREDISPOSED TO THROMBOGENICITY. DEVELOPMENT OF THROMBUS IS A COMMON COMPLICATION IS SUCH PATIENT POPULATION. CRITICALLY ILL PATIENTS ARE AT INCREASED RISK OF VTE BECAUSE OF THE PRESENCE OF MULTIPLE PREDISPOSING FACTORS [W. GEERTS, ET AL. VENOUS THROMBOEMBOLISM AND ITS PREVENTION IN CRITICAL CARE. J CRIT CARE, 17 (2002), PP. 95-104]. THE RATE OF THROMBOSIS FOR CRITICAL CARE PATIENTS RECEIVING CVCS RANGES FROM 20 TO 30%, PATIENTS WITH PERIPHERAL CENTRAL CATHETERS HAD A SIGNIFICANTLY HIGHER INCIDENCE RATE OF DVT THAN PATIENTS WITH CVC (27.2% VS 9.6%, P=0.0012). THE RATE OF DVT IN THE PATIENT POPULATION RECEIVING IVTM POST-CARDIAC ARREST IS 1.7%. FOUR RANDOMIZED CONTROLLED CLINICAL TRIALS CONDUCTED IN A TOTAL OF 943 PATIENTS SHOWED THAT THERE WAS NO DIFFERENCE IN THE DVT RATE WHEN COMPARING ZOLL IVTM CATHETERS TO STANDARD CVCS. THE RATE OF DVT IN THE PATIENT POPULATION RECEIVING SURFACE COOLING HAS BEEN REPORTED BETWEEN 3 AND 15%. TIMELY ADMINISTRATION OF PROPHYLACTIC ANTICOAGULATION IS SAFE AND SIGNIFICANTLY REDUCES DVT RATES IN HIGH RISK PATIENT POPULATIONS [ZOLL WHITE PAPER ON DVT]. DVT IS ANTICIPATED EVENT AND COULD POTENTIALLY OCCUR WITH USAGE OF ANY CATHETER. THE REPORTED EVENT IS PROBABLY RELATED TO THE DEVICE AND NOT RELATED TO THE PROCEDURE.
AS REPORTED, AFTER REMOVING THE ICY CATHETER (LOT #UNKNOWN) FROM A 20-YEAR-OLD MALE PATIENT WHO WAS TRANSPORTED AFTER HANGING. D-DIMER INCREASED, AND THROMBOSIS WAS FOUND FROM CATHETER INSERTION SITE TO RENAL VEIN INFLOW BY CT SCAN. HEPARIN INJECTION AND THROMBOLYSIS BEGAN. PER REPORTER, THE ICY CATHETER INSERTION WAS SMOOTH INTO THE RIGHT FEMORAL VEIN BY A EXPERIENCED PHYSICIAN WITH ZOLL PRODUCTS. THE PATIENT IS ALIVE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 978366 | ZOLL IVTM ICY CATHETER | CENTRAL VENOUS CATHETER | NCX | ZOLL CIRCULATION | 8700-0782-03 | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 20 YR | Male | Other |