AUGMENT INJECTABLE KIT 1.5CC INJECTABLE USA
Report
- Report Number
- 3003897776-2023-00001
- Event Type
- Malfunction
- Date Received
- January 18, 2023
- Date of Event
- December 21, 2022
- Report Date
- April 7, 2023
- Manufacturer
- BIOMIMETIC THERAPEUTICS, LLC
- Product Code
- NOX
- UDI-DI
- 10859477002133
- PMA / PMN Number
- P100006
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
DEVICE WILL NOT BE RETURNED. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, IT WILL BE PROVIDED ON A SUPPLEMENTAL REPORT.
THE REPORTED EVENT COULD NOT BE CONFIRMED (PACKAGING - MISSING PRODUCT / COMPONENT) SINCE THE DEVICE WAS NOT RETURNED FOR EVALUATION AND NO OTHER EVIDENCES WERE PROVIDED. QUALITY IN ARLINGTON EXAMINED UNITS FROM THE SAME LOT AND DETERMINED ALL COMPONENTS WERE PRESENT. IT WAS ALSO NOTED THAT NO OTHER COMPLAINTS HAVE BEEN RECEIVED FOR THIS LOT INDICATING A MISSING COMPONENT. MORE DETAILED INFORMATION ABOUT THE COMPLAINT EVENT AS WELL AS THE AFFECTED DEVICE MUST BE AVAILABLE IN ORDER TO DETERMINE THE ROOT CAUSE OF THE COMPLAINT EVENT. A REVIEW OF THE DEVICE HISTORY FOR THE REPORTED LOT DID NOT INDICATE ANY ABNORMALITIES. NO CORRECTIVE ACTIONS ARE REQUIRED AT THIS TIME. A REVIEW OF THE LABELING DID NOT INDICATE ANY ABNORMALITIES. NO INDICATIONS OF MATERIAL, MANUFACTURING OR DESIGN RELATED PROBLEMS WERE FOUND DURING THE INVESTIGATION. IF MORE INFORMATION IS PROVIDED, THE CASE WILL BE REASSESSED.
IT WAS REPORTED THAT THE 1.5 CC LIQUID VIAL WAS MISSING FROM THE KIT. ANOTHER COMPANIES INJECTABLE WAS USED TO COMPLETE THE CASE.
IT WAS REPORTED THAT THE 1.5 CC LIQUID VIAL WAS MISSING FROM THE KIT. ANOTHER COMPANIES INJECTABLE WAS USED TO COMPLETE THE CASE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 363285 | AUGMENT INJECTABLE KIT 1.5CC INJECTABLE USA | FILLER, BONE VOID, SYNTHETIC PEPTIDE | NOX | BIOMIMETIC THERAPEUTICS, LLC | K30001510 | 1737280 | 10859477002133 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |