BALL HEADS: MECTACER BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 36 SIZE M
Report
- Report Number
- 3005180920-2022-00993
- Event Type
- Injury
- Date Received
- January 18, 2023
- Date of Event
- December 16, 2022
- Report Date
- January 18, 2023
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LZO
- UDI-DI
- 07630030809200
- PMA / PMN Number
- K112115
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- 003
Narratives
BATCH REVIEW PERFORMED ON 22 DECEMBER 2022 LOT 2207392: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 05-JULY-2022. EXPIRATION DATE: 2027-06-21. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ADDITIONAL IMPLANTS INVOLVED: BATCH REVIEW PERFORMED ON 22 DECEMBER 2022 ON LINER: MPACT 01.32.3644HCAT HOODED PE HC LINER Ø36/E (K132879) LOT. 2201678. LOT 2201678: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 03-MARCH-2022. EXPIRATION DATE: 2027-02-17. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. BATCH REVIEW PERFORMED ON 22 DECEMBER 2022 ON STEM: M-VIZION 01.22.105 DISTAL STEM Ø16MM L 140MM STRAIGHT (K170690) LOT. 2010764. LOT 2010764: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 16-JUNE-2021. EXPIRATION DATE: 2026-06-06. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH ANOTHER SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. BATCH REVIEW PERFORMED ON 22 DECEMBER 2022 ON STEM: M-VIZION 01.22.404 PROXIMAL BODY Ø20MM L 70MM STD WITH HOLES (K201471) LOT. 2010777. LOT 2010777: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 04-NOV-2021. EXPIRATION DATE: 2026-10-24. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. BATCH REVIEW PERFORMED ON 22 DECEMBER 2022 ON CUP: MPACT 01.32.152DH ACETABULAR SHELL Ø52 TWO-HOLES (K132879) LOT. 2115587. LOT 2115587: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 19-JAN-2022. EXPIRATION DATE: 2027-01-09. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW.
THE PATIENT HAD A PRIMARY HIP SURGERY ON (B)(6) 2022. ON (B)(6) 2022, THE PATIENT CAME IN REPORTING PAIN DUE TO A PERIPROSTHETIC FRACTURE THAT WAS CAUSED FROM FALLING. THE SURGEON CABLED THE FRACTURE, AND REVISED THE STEM, HEAD, AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY. ON (B)(6) 2022, THE PATIENT CAME IN DUE TO SIGNS OF INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PREFORMED A WASHOUT AND REVISED THE HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY. PRESENTLY, ON (B)(6) 2022, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED ALL IMPLANTS AND THE SURGERY WAS COMPLETED SUCCESSFULLY. THE INFECTION CASE OF THE (B)(6) 2022 WAS REPORTED IN THE MDR 2022-00976, IN WHICH IT IS NOT MENTIONED THE PERIPROSTHETIC FRACTURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 3773 | BALL HEADS: MECTACER BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 36 SIZE M | HIP CERAMIC BALL HEAD | LZO | MEDACTA INTERNATIONAL SA | 01.29.209 | 2207392 | 07630030809200 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Male | Required Intervention |