AUTOFEED HUMIDIFICATION CHAMBER
Report
- Report Number
- 9611451-2010-00092
- Event Type
- Malfunction
- Date Received
- February 24, 2010
- Date of Event
- January 25, 2010
- Report Date
- January 25, 2010
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- PMA / PMN Number
- K934140
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NZ
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE RETURNED CHAMBER WAS VISUALLY INSPECTED FOR EVIDENCE OF A DETACHED SPIKE AND/OR BROKEN TUBE. RESULTS: OUR EXAMINATION OF THE DEVICE CONFIRMED THAT THE VENT SPIKE HAD PARTED FROM THE WATER FEEDSET TUBING, JUST SHORT OF WHERE BOTH COMPONENTS CONNECT TO EACH OTHER. THE RIM OF THE TUBING APPEARED TO BE ROUGH WITH VISIBLE TRACK MARKS, CONSISTENT WITH A TEAR OR PULL. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE, FOR ANY COUNTRY WORLDWIDE, FOR THE LOT NUMBER PROVIDED (091126). CONCLUSION/COMMENTS: IT IS LIKELY THAT THE WATER FEEDSET TUBING BROKE AWAY FROM THE VENT SPIKE AS A RESULT OF SUBSTANTIAL FORCE APPLIED TO IT, FOR INSTANCE IF ACCIDENTALLY PULLED OR STRETCHED. THIS IS CONSISTENT WITH THE HOSPITAL'S REPORT THAT THE DAMAGE OCCURRED AT THE TIME THE WATERBAG WAS BEING CHANGED. AS PART OF OUR MANUFACTURING PROCESS, EVERY MR290V CHAMBER UNDERGOES PRESSURE TESTING FOR POTENTIAL LEAKS. HAD THE FEEDSET TUBING BEEN DAMAGED IN SUCH A MANNER AT THE TIME OF PRODUCTION, THE CHAMBER WOULD NOT HAVE PASSED THE FINAL PRESSURE TEST AND BEEN REJECTED. OUR USER INSTRUCTIONS WHICH ACCOMPANIES THE MR290 CHAMBER SPECIFIES THE IMPORTANCE OF PERFORMING A PRESSURE/LEAK TEST ON THE SYSTEM PRIOR TO PATIENT CONNECTION. THE INSTRUCTIONS ALSO REMIND THE USER TO SET THE APPROPRIATE VENTILATOR ALARMS TO ALERT ATTENDING HOSPITAL STAFF IN THE EVENT THAT A LEAK IN THE SYSTEM DEVELOPS DURING USE. (B)(4).
A HOSPITAL IN (B)(6) ('THE HOSPITAL') REPORTED TO A FISHER & PAYKEL HEALTHCARE ('FPH') FIELD REPRESENTATIVE THAT THE WATERFEED SET TUBING OF AN MR290V AUTOFEED HUMIDIFICATION CHAMBER ('THE CHAMBER') BROKE OFF IN THE REGION WHERE IT CONNECTED TO THE CHAMBER'S VENT SPIKE. THE HOSPITAL FURTHER ADVISED FPH THAT THE BREAKAGE OCCURRED AT THE TIME THE WATERBAG WAS BEING CHANGED, PROBABLY AS THE VENT SPIKE WAS BEING PULLED FROM THE WATER BAG. THE HOSPITAL ALSO CONFIRMED THAT NEITHER THE PATIENT NOR ATTEMPTING HOSPITAL STAFF AND/OR CAREGIVER(S) SUSTAINED ANY INJURY OR OTHER CONSEQUENCE AS A RESULT OF THIS INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AUTOFEED HUMIDIFICATION CHAMBER | BTT | FISHER & PAYKEL HEALTHCARE LTD | MR290V | 091126 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |