FDA Adverse Event Malfunction Summary report: N

MI-750

MDR report key: 16184645 · Received January 17, 2023

Report

Report Number
2028295-2023-00001
Event Type
Malfunction
Date Received
January 17, 2023
Date of Event
December 16, 2022
Report Date
January 12, 2023
Manufacturer
MEDICAL ILLUMINATION INTRNATION INC.
Product Code
FSY
UDI-DI
00816835020201
PMA / PMN Number
K140812
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FIELD FAILURE (CCR #2754) OF AN MI-LINE, SINGLE ARM, MI-750 LIGHT HEAD SYSTEM (MODEL: 061424) WAS RECEIVED (01/12/23) INTO ENGINEERING FOR FAILURE ANALYSIS. THE CUSTOMER STATES THAT THE UNIT FELL FROM THE CEILING MOUNT. BASED ON THE CUSTOMER SUPPLIED INFORMATION, THE EXTENSION ARM (1000724-S) WITH SPRING ARM (1000727-MI750) IS HANGING FROM THE POWER CABLE (1000789). OBSERVATIONS: FROM THE CUSTOMER SUPPLIED INFORMATION. THE SUPPORTING SNAP RING (0003445) AND ITS SUPPORT WASHER (0003443) WHICH WOULD HAVE BEEN SITTING IN THE SNAP RING GROOVE ON THE END OF THE SHAFT (1001462) WAS NOT PRESENT AS EXPECTED. IF THE SNAP RING HAD DISLODGED ITSELF FROM THE GROOVE, IT SHOULD HAVE BEEN FOUND IN THE 1001378 HUB COVER ALONG WITH THE SUPPORT WASHER ON THE FAR SIDE OF THE BEARING PLATE (1001464). THE CONDITION AS RECEIVED (OUT OF THE BOX) WE FOUND THE SNAP RING & THE SUPPORT WASHER HAD BEEN INSTALLED BACK ONTO THE SNAP RING GROOVE ON THE END OF THE SHAFT. THE HUB COVER WAS IN PLACE AND MOUNTED ON THE END OF THE EXTENSION ARM HOUSING. IT IS NOT POSSIBLE FOR THE SNAP RING AND SUPPORT WASHER TO GET PAST THE INTERNAL BEARING PLATE WHICH IS HELD IN PLACE IN A COUNTER BORE IN THE EXTENSION ARM HOUSING. RESTORING THE ASSEMBLY TO ITS NORMAL CONDITION: ENGINEERING RE-BUILT THE ASSEMBLY WITH THE SNAP RING & SUPPORT WASHER. WE ASSURED THAT THE SNAP RING WAS SET PROPERLY IN ITS GROOVE AND APPLIED FORCE TO SEE IF THE EXTENSION ARM HELD ITSELF IN PLACE. WE EXAMINED THE SNAP RING & THE SNAP RING GROOVE FOR PROPER FIT AND DAMAGE. NO DAMAGE FOUND. CONCLUSION: THERE WASN'T ENOUGH OR SUFFICIENT DETAILED INFORMATION RECEIVED FROM THE CUSTOMER TO INDICATE HOW THIS FAILURE COULD HAVE OCCURRED OR ESTABLISH A ROOT CAUSE. IF THERE HAD BEEN AN ASSEMBLY ERROR, THE AS RECEIVED EVIDENCE DID NOT INDICATE IT. THE AS-RECEIVED, OUT-OF-THE-BOX EVIDENCE PROVIDED MAY HAVE BEEN ALTERED MAKING THE FAILURE ANALYSIS DIFFICULT. THERE ISN'T ANY PRIOR INCIDENT OF THIS PARTICULAR FAILURE MODE. IF ADDITIONAL INFORMATION CAN BE UPTAINED FROM THE END USER, WILL UPDATE REPORT.

Description of Event or Problem · 0

ON DECEMBER 16TH, IT WAS REPORTED TO MEDICAL ILLUMINATION THAT AN MI750 LIGHT SYSTEM HAS SEPARATED FROM THE DOWN TUBE. NO INJURIES OCCURED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1223658 MI-750 MINOR SURGICAL LIGHT FSY MEDICAL ILLUMINATION INTRNATION INC. 061424 00816835020201

Patients

Seq Age Sex Outcome Treatment
1 Unknown