PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM

MDR report key: 16184510 · Received January 17, 2023

Report

Report Number
2024168-2023-00593
Event Type
Injury
Date Received
January 17, 2023
Date of Event
December 26, 2022
Report Date
February 14, 2023
Manufacturer
ABBOTT VASCULAR
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION. THE ADDITIONAL PROSTYLE DEVICES REFERENCED ARE BEING FILED UNDER SEPARATE MEDWATCH REPORT NUMBERS.

Additional Manufacturer Narrative · 0

ANALYSIS WAS PERFORMED ON THE RETURNED DEVICE. THE REPORTED NEEDLE TO CUFF MISS COULD NOT BE TESTED/ CONFIRMED AS THE PLUNGER, LINK, SUTURE, POSTERIOR NEEDLE TIP AND CUFFS WERE NOT RETURNED. A REVIEW OF THE LOT HISTORY RECORD IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. THE REPORTED DIFFICULTY AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR INABILITY TO MAINTAIN POSITION/STABILITY OF THE DEVICE DURING DEPLOYMENT DUE TO CIRCUMSTANCES OF THE PROCEDURE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING. D4: LOT NUMBER UPDATED FROM 2072941 TO 2090841.

Description of Event or Problem · 0

IT WAS REPORTED THAT THIS WAS AN ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY USING A PROSTYLE AFTER A NEURO INTERVENTION (ANEURYSM) PROCEDURE USING A 6F SHEATH. REPORTEDLY, A CUFF MISS OCCURRED. ANOTHER PROSTYLE WAS ATTEMPTED BUT THE SAME OCCURRED. ANOTHER PROSTYLE WAS ATTEMPTED AGAIN BUT THE SAME OCCURRED. MANUAL COMPRESSION WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
818795 PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM SUTURE MEDIATED CLOSURE MGB ABBOTT VASCULAR 2090841

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention