FDA Adverse Event Malfunction Summary report: N

OMNIPOD DASH¿ INSULIN MANAGEMENT SYSTEM

MDR report key: 16179479 · Received January 16, 2023

Report

Report Number
3004464228-2023-01229
Event Type
Malfunction
Date Received
January 16, 2023
Date of Event
December 23, 2022
Report Date
January 7, 2023
Manufacturer
INSULET CORPORATION
Product Code
LZG
UDI-DI
10385082000139
PMA / PMN Number
K192659
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

A TECHNICAL ASSESSMENT OF THE DEVICE DESIGN IDENTIFIED THE ROOT CAUSE OF THE THERMAL EVENT TO BE THE DASH PDM CHARGING VOLTAGE EXCEEDING THE BATTERY SPECIFICATION, DEFINED AS OVERCHARGING. FIELD SAFETY CORRECTIVE ACTION HAS BEEN INITIATED BY INSULET CORPORATION.

Description of Event or Problem · 0

IT WAS REPORTED BY THE PATIENT THAT THE OMNIPOD PERSONAL DIABETES MANAGER (PDM) BATTERY GOT VERY HOT WHILE CHARGING OVERNIGHT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1657339 OMNIPOD DASH¿ INSULIN MANAGEMENT SYSTEM PUMP, INFUSION, INSULIN LZG INSULET CORPORATION PT-000010 10385082000139

Patients

Seq Age Sex Outcome Treatment
1 32 YR Female