FDA Adverse Event Malfunction Summary report: N

ALINITY C PROCESSING MODULE

MDR report key: 16174435 · Received January 16, 2023

Report

Report Number
3016438761-2023-00022
Event Type
Malfunction
Date Received
January 16, 2023
Report Date
February 13, 2023
Manufacturer
ABBOTT LABORATORIES
Product Code
JJE
UDI-DI
00380740137380
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DURING THE REQUESTED SITE VISIT, THE FIELD SERVICE REPRESENTATIVE (FSR) INSPECTED THE ANALYZER AND REPLACEMENT OF THE TUBING, PERISTALTIC HEAD (ROHS)_PART NUMBER 7-202464-0 LEAKING FROM THE CUVETTE WASH. THE REPLACEMENT OF PART TUBING, PERISTALTIC HEAD (ROHS) (7-202464-01), WHICH RESOLED THE ISSUE. A REVIEW OF THE ALINITY CI PROCESSING MODULE, SERIAL NUMBER (B)(6) SERVICE HISTORY WAS PERFORMED AND NO ADDITIONAL ERRATIC RESULTS PRE- OR POST THE CURRENT COMPLAINT WERE IDENTIFIED. A REVIEW OF TRACKING AND TRENDING OF THE ALINITY C PROCESSING MODULE DID NOT IDENTIFY ANY TRENDS ASSOCIATED WITH THE COMPLAINT ISSUE. A REVIEW OF TRACKING AND TRENDING DID NOT IDENTIFY ANY TRENDS FOR THE TUBING, PERISTALTIC HEAD (ROHS). A REVIEW OF HISTORICAL DATA REVEALED NO TRENDS, SYSTEMIC ISSUES, OR RELATED NONCONFORMANCES. THE ALINITY CI-SERIES OPERATIONS MANUAL AND ALINITY C SERVICE DOCUMENTATION PROVIDE ADEQUATE INFORMATION REGARDING THE TROUBLESHOOTING OF ERRATIC/DISCREPANT RESULTS AND THE REMOVAL, REPLACEMENT AND VERIFICATION OF PART NUMBER 7-202464-01 TUBING, PERISTALTIC HEAD (ROHS). A REVIEW OF THE PRODUCT LABELING CONCLUDED THAT THE ISSUE IS SUFFICIENTLY ADDRESSED. BASED ON THE INVESTIGATION NO PRODUCT DEFICIENCY WAS IDENTIFIED FOR EITHER THE ALINITY CI PROCESSING MODULE, SERIAL NUMBER (B)(6), OR THE TUBING, PERISTALTIC HEAD (ROHS).

Additional Manufacturer Narrative · 0

COMPLETE INFORMATION FOR SECTION A PATIENT INFORMATION, 1. PATIENT IDENTIFIER = SID= (B)(6) AND (B)(6). AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSELY DECREASED SODIUM RESULTS ON ALINITY C PROCESSING MODULE FOR MULTIPLE PATIENTS. THE RESULTS PROVIDED WERE: SID (B)(6) INITIAL= 114 MMOL/L /REPEATED =137 MMOL/L. SID (B)(6) INITIAL=120 MMOL/L /REPEATED= 140 MMOL/L. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
819186 ALINITY C PROCESSING MODULE ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE ABBOTT LABORATORIES 03R6701 00380740137380

Patients

Seq Age Sex Outcome Treatment
1 Unknown