FDA Adverse Event
Other
Summary report: N
MICROSELECTRON-HDR
MDR report key: 161684
·
Received April 2, 1998
Report
- Report Number
- 1121753-1998-00009
- Event Type
- Other
- Date Received
- April 2, 1998
- Date of Event
- July 23, 1997
- Report Date
- March 5, 1998
- Manufacturer
- NUCLETRON BV
- Product Code
- JAQ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE SCREW MECHANISM BROKE OFF OF LOCKING PIPE, WHICH APPEARED TO RESULT FROM A SOLDER JOINT FAILURE. NO SIGN OF CORROSION OF EXCESSIVE FORCE COULD BE NOTED UPON VISUAL INSPECTION OF THE FAILED PART, ITEM WAS REPLACED BY NUCLETRON. PART IS BEING RETURNED TO NUCLETRON US UNDER RMA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MICROSELECTRON-HDR | REMOTE CONTROLLED RADIONUCLIDE APPLICATOR SYSTEM | JAQ | NUCLETRON BV | M/HDR (ACCESSORY) | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Other |