3085 SURGICAL TABLE
Report
- Report Number
- 1043572-2010-00006
- Date Received
- February 26, 2010
- Date of Event
- January 29, 2010
- Report Date
- February 26, 2010
- Manufacturer
- STERIS CORPORATION - MONTGOMERY
- Product Code
- FQO
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
A STERIS SERVICE TECHNICIAN INSPECTED THE TABLE AND WAS UNABLE TO DUPLICATE THE REPORTED EVENT. HOWEVER THE TECHNICIAN DID FIND THE POWER CORD COVERING WAS TORN, EXPOSING ELECTRICAL WIRES, DUE TO THE SURGICAL TABLE BEING REPEATEDLY ROLLED OVER THE CORD. IN ADDITION THE TECHNICIAN FOUND THE TABLE COLUMN COVER ATTACHING HOOKS WERE BENT, WHICH CAUSED THE COVERS TO COLLAPSE AND POSSIBLY INTERFERE WITH THE ELECTRICAL COMPONENTS IN THE TABLE COLUMN. UPON VISUAL INSPECTION DURING A SUBSEQUENT VISIT THE TECHNICIAN DID FIND THE COLUMN SHROUD WAS DAMAGED. WARNING LABELS WERE PRESENT ON THE TABLE, INCLUDING "POSSIBLE TABLE DAMAGE - DO NOT USE THE TABLE BASE FOR STORAGE". THE CAUSE OF THIS EVENT APPEARS TO BE MISUSE OF THE DEVICE - I.E., SETTING/STORING OBJECTS ON THE BASE OF THE TABLE - RESULTING IN DAMAGE TO THE TABLE'S ELECTRONICS, WHICH IN EXTREME CASES CAN CAUSE INADVERTENT MOVEMENT OF THE TABLE. SUCH MISUSE IS SPECIFICALLY CONTRADICTED IN THE LABELING OF AMSCO 3085SP SURGICAL TABLES AND IN STERIS TRAINING REGARDING THE DEVICE. THIS TYPE OF MISUSE IS THE OBJECT OF A VOLUNTARY FIELD CORRECTION INITIATED BY STERIS CORPORATION ON (B)(4) 2010 (REPORT # 1043572-2/17/10-003C) OF 3085 SP SURGICAL TABLES. AS PART OF THE FIELD CORRECTION, STERIS DEVELOPED AND IS INSTALLING A RIGID GUARD ON ALL TABLES THAT PROTECTS THE ELECTRONIC SWITCH ASSEMBLY FROM CONTACT WITH TABLE COMPONENTS THAT CAN BECOME DAMAGED AND IMPINGE THE SWITCH ASSEMBLY AS A RESULT OF ITEMS BEING IMPROPERLY SET OR STORED ON THE BASE OF THE TABLE. THE SURGICAL TABLE IS NOT UNDER STERIS SERVICE CONTRACT AND IS CURRENTLY MAINTAINED AND SERVICED BY THE FACILITY BIOMEDICAL DEPARTMENT. THE STERIS SERVICE TECHNICIAN INSTALLED THE RIGID GUARD AND BIOMEDICAL PERSONNEL REPAIRED THE TABLE AFTER THE INCIDENT, PLACING THE TABLE BACK IN SERVICE. NO FURTHER ISSUES HAVE BEEN REPORTED WITH THE SURGICAL TABLE SINCE THE REPORTED EVENT STERIS HAS OFFERED TO PROVIDE AN IN-SERVICE TRAINING TO HOSPITAL STAFF REGARDING THE PROPER USE AND OPERATION OF THE TABLE.
THE USER FACILITY REPORTED THAT AT THE COMPLETION OF A PATIENT PROCEDURE THE SURGICAL TABLE MOVED UPWARD WHEN IT WAS COMMANDED TO MOVE DOWNWARD. ALSO, ONE OF THE BASE COVERS CAME OFF CAUSING THE OTHER BASE COVERS TO COLLAPSE. NO INJURY WAS REPORTED TO EITHER THE PATIENT OR THE HOSPITAL STAFF.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 3085 SURGICAL TABLE | SURGICAL TABLE | FQO | STERIS CORPORATION - MONTGOMERY | 3085SP |