FDA Adverse Event Malfunction Summary report: N

BD SMARTSITE¿ NEEDLE-FREE CONNECTOR

MDR report key: 16164150 · Received January 13, 2023

Report

Report Number
9616066-2022-02211
Event Type
Malfunction
Date Received
January 13, 2023
Date of Event
November 30, 2022
Report Date
December 27, 2022
Manufacturer
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
Product Code
FPA
PMA / PMN Number
K960280
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

2000E CHINA 510K: THIS IS AN INTERNATIONAL CODE - THE MODEL#:/CATALOG#: IDENTIFIED IS A CAREFUSION PRODUCT WHICH IS SAME OR SIMILAR TO A DEVICE THAT IS APPROVED FOR SALE DOMESTICALLY. THE DOMESTIC SIMILAR LIST NUMBER IS 2000E. THE 510K NUMBER PROVIDED IS FOR THE DOMESTIC SIMILAR PRODUCT: K960280. INVESTIGATION SUMMARY: A 2000E CHINA PRODUCT WAS NOT AVAILABLE FOR INVESTIGATION; HOWEVER, THE CUSTOMER INDICATED THE COMPLAINT SAMPLE WAS FROM LOT: 22026149. THE FEEDBACK PROVIDED BY THE CUSTOMER SUGGESTS A COMPLETE OCCLUSION WAS DETECTED AT THE SMARTSITE DEVICE. NO FURTHER INFORMATION WAS AVAILABLE TO ASSIST THE INVESTIGATION IN THIS INSTANCE. THE DETAILS OF THIS FEEDBACK WERE FORWARDED TO THE MANUFACTURING SITE FOR INVESTIGATION. A REVIEW OF THE PRODUCTION RECORDS FOR LOT: 22026149 DID NOT IDENTIFY ANY IN-PROCESS TESTING FAILURES OR QUALITY DEVIATIONS WHICH MAY HAVE RESULTED IN A REPORT OF THIS NATURE. THE ROOT CAUSE OF THE CUSTOMER¿S EXPERIENCE COULD NOT BE DETERMINED AS THE SAMPLE WAS NOT AVAILABLE FOR INVESTIGATION. IN THIS INSTANCE, WITHOUT A SAMPLE, IT IS NOT POSSIBLE TO DETERMINE WHETHER A MANUFACTURING DEFECT COULD HAVE CAUSED OR CONTRIBUTED TO THE CUSTOMER¿S EXPERIENCE. PREVIOUS COMPLAINTS FOR OCCLUSIONS HAVE BEEN RELATED TO FEATURES ON THE SURFACE OF THE MALE LUER OF THE CONNECTING PRODUCTS. THESE FEATURES INCLUDE FLASH OR A RAISED EDGE TO THE TIP OF THE MALE LUER WHICH HAVE PREVIOUSLY BEEN SHOWN TO INTERMITTENTLY LEAD TO RESTRICTED FLOW DUE TO THEM PINCHING THE BLUE PISTON OF THE SMARTSITE® AND NOT ALLOWING IT TO OPEN. THIS CAN SOMETIMES BE RESOLVED BY DISCONNECTING AND REATTACHING THE SAME LUER CONNECTION WHICH MAY REPOSITION THE LUER AGAINST THE PISTON AND IMPROVE THE FLOW, OR ALTERNATIVELY BY CHANGING THE CONNECTING MALE LUER. THE CONNECTING PRODUCT IN CLINICAL USE AT THE TIME OF THIS SPECIFIC CUSTOMER'S EXPERIENCE WAS NOT RECEIVED FOR INVESTIGATION; THEREFORE, IT WAS NOT POSSIBLE TO CONFIRM IF THE CONNECTING PRODUCT MAY HAVE CONTRIBUTED TO THE REPORTED ISSUE. A REVIEW OF THE CUSTOMER FEEDBACK DATABASE INDICATES THAT THIS IS A RARE OCCURRENCE WITH A SMALL NUMBER OF SIMILAR REPORTS AGAINST THE SMARTSITE COMPONENT IN THE PAST 12 MONTHS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD SMARTSITE¿ NEEDLE-FREE CONNECTOR INFUSION CHANNEL WAS BLOCKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE: "IT IS FOUND THAT THE ORIGINAL SMOOTH INFUSION CHANNEL IS BLOCKED AFTER A FEW NEEDLELESS AIRTIGHT INFUSION NEEDLES ARE CONNECTED AND USED. OTHERS: NO PATIENTS INVOLVED".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2074760 BD SMARTSITE¿ NEEDLE-FREE CONNECTOR INTRAVASCULAR ADMINISTRATION SET FPA SISTEMAS MEDICOS ALARIS, S.A. DE C.V. 22026149

Patients

Seq Age Sex Outcome Treatment
1 Unknown