FDA Adverse Event Malfunction Summary report: N

PROLIEVE THERMODILITATION SYSTEM

MDR report key: 1616284 · Received February 26, 2010

Report

Report Number
3005099803-2010-00698
Event Type
Malfunction
Date Received
February 26, 2010
Date of Event
June 2, 2009
Report Date
February 4, 2010
Manufacturer
BOSTON SCIENTIFIC - MARLBOROUGH
Product Code
MEQ
PMA / PMN Number
P030006
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE SUSPECT DEVICE, A PROLIEVE THERMODILATATION SYSTEM, REFURBISHED, WAS SENT TO SANMINA FOR ANALYSIS. EVALUATION RESULTS STATE THE PRESSURE FAILED DUE TO A DEFECTIVE SENSOR BOARD AND RADIO FREQUENCY (RF) OUT OF SPECIFICATION WAS DUE TO A DEFECTIVE Q1 ON RF CONTROL BOARD. THE MALFUNCTIONS WERE ADDRESSED BY REPLACING THE SENSOR BOARD, RF OUTPUT MODULE AND RF CONTROL BOARD. IN ADDITION, THE FOLLOWING WERE ALSO REPLACED: DEFECTIVE PELTIERS IN THE HEAT EXCHANGER MODULE FAN, BROKEN FAN ON HEAT EXCHANGER MODULE AND DEFECTIVE PHYSITEMPS 1 & 3.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PROLIEVE THERMODILATATION SYSTEM, REFURBISHED, WAS USED DURING A BENIGN PROSTATIC HYPERPLASIA (BPH) PROCEDURE. A BSC SALES REPRESENTATIVE CALLED INTO TECHNICAL SERVICES DURING THE TREATMENT TO REPORT A PROBLEM WITH PROLIEVE UNIT; THERE WAS NOT ENOUGH PRESSURE DURING THE TREATMENT. THE SYSTEM PUMP SPEED WAS UP TO 50 BEFORE IS REACHED ~10 PSI. THE CASE WAS COMPLETED WITH NO PATIENT ISSUES. PER FOLLOW UP WITH THE COMPLAINANT, THE PATIENT CONDITION IS GOOD. NOTE: THE EVENT, AS REPORTED, DID NOT REFLECT AN MDR-REPORTABLE SCENARIO. HOWEVER, EVALUATION OF THE RETURNED CONSOLE REVEALED AN MDR-REPORTABLE MALFUNCTION OF "RADIO FREQUENCY (RF) OUT OF SPECIFICATION". THE MDR IS BASED UPON THE EVALUATION RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROLIEVE THERMODILITATION SYSTEM SYSTEM, HYPERTHERMIA, RF/MICROWAVE (BENIGN PROSTATIC HYPERPLASIA),THERMOTHERAPY MEQ BOSTON SCIENTIFIC - MARLBOROUGH M006880806R0

Patients

Seq Age Sex Outcome Treatment
1 70 YR