DA VINCI SI SURGICAL SYSTEM
Report
- Report Number
- 2955842-2010-00068
- Event Type
- Other
- Date Received
- February 25, 2010
- Date of Event
- January 26, 2010
- Report Date
- January 26, 2010
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K081207
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY
- Reporter Occupation
- OTHER
Narratives
THE INVESTIGATION CONDUCTED BY FIELD SERVICE ENGINEERING CONCLUDED THAT SYSTEM ERROR CODES 41049 AND 23017 WERE ASSOCIATED WITH THE REMOTE ARM CONTROLLER (RAC) AND/OR THE MASTER TOOL MANIPULATOR (MTM) GIMBAL. THE RAC CONSISTS OF FIVE PRINTED CIRCUIT ASSEMBLY (PCA) BOARDS WHICH OPERATE TOGETHER TO PROVIDE CONTROL OF THE SYSTEM ARMS. THE MTM GIMBAL IS A SUBSECTION OF THE COMPLETE MTM ARM, CONSISTING OF THE LINKS ASSOCIATED WITH THE FIVE AXES OF MOTION CLOSEST TO THE SURGEON'S HAND. THESE AXES MEASURE THE ORIENTATION OF THE MTM HANDLE AND THE MTM GRIP ANGLE. THE GIMBAL IS SEPARATED FROM THE REST OF THE MTM BY REMOVING THE PLATFORM LINK FROM THE FOREARM LINK. THE SYSTEM WAS REPAIRED BY REPLACING THE AFFECTED RAC AND MTM GIMBAL. THE SYSTEM ALARM (SYSTEM GENERATED FAULT CODE) FUNCTIONED AS DESIGNED AND THERE WAS NO INJURY TO THE PATIENT. THE 41049 SYSTEM ERROR CODE APPEARS WHEN THE DA VINCI SI SAFETY SYSTEM (SUPERVISOR) DETECTED AN INTERNAL LOGIC ERROR, THUS PUTTING THE SYSTEM IN A RECOVERABLE SAFE STATE. SYSTEM ERROR CODE 23017 APPEARS WHEN THE DA VINCI S SAFETY SYSTEM DETERMINES THAT A MOTOR DID NOT RESPOND AS EXPECTED AND THE MEASURED MOTION DID NOT MATCH THE INTERNAL SIMULATION OF THE MOTOR. UPON DETERMINING THIS CONDITION, THE SAFETY SYSTEMS PUT DA VINCI IN A RECOVERABLE SAFE STATE. THE RAC AND MTM GIMBAL WERE RETURNED TO ISI FOR EVALUATION. ENGINEERING DETERMINED THAT THE RAC DRIVE MODULE WAS DEFECTIVE, RESULTING IN THE CUSTOMER EXPERIENCED SYSTEM ERROR CODES. THE MTM GIMBAL WAS EVALUATED AND FOUND TO BE WITHIN SPECIFICATIONS WITH NO DEFECTS. AS OF FEBRUARY 24, 2010, THERE HAVE BEEN NO REPORTED RECURRENCES OF THE ISSUE AT THIS HOSPITAL.
IT WAS REPORTED THAT DURING A DA VINCI SI SURGICAL PROCEDURE, THE SITE EXPERIENCED SYSTEM ERROR CODE 41049 WHILE CLUTCHING A PATIENT SIDE MANIPULATOR (PSM) ARM. THE SYSTEM WAS ALSO FAULTING WITH MULTIPLE OCCURRENCES OF SYSTEM ERROR CODE 23017. THE SURGICAL STAFF MADE THE DECISION TO NOT TROUBLESHOOT THE SYSTEM ERRORS AND ABORTED THE PROCEDURE AFTER ANESTHESIA HAD BEEN ADMINISTERED AND PORTS PLACED. NO PATIENT HARM WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DA VINCI SI SURGICAL SYSTEM | ENDOSCOPIC INSTRUMENT CONTROL SYSTEM | NAY | INTUITIVE SURGICAL,INC. | IS3000 A6.1P8 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | DA VINCI SI INSTRUMENTS, ACCESSORIES, AND ESU |