FDA Adverse Event Other Summary report: N

MICROSELECTRON-HDR (CLASSIC)

MDR report key: 161556 · Received April 2, 1998

Report

Report Number
1121753-1998-00005
Event Type
Other
Date Received
April 2, 1998
Date of Event
November 22, 1997
Report Date
February 20, 1998
Manufacturer
NUCLETRON BV
Product Code
JAQ
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

QUALITY ASSURANCE WITH NO PT PRESENT REVEALED THAT IN THE ABSENCE OF AN APPLICATOR IN CHANNEL 18, THE SOURCE LEFT THE SAFE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MICROSELECTRON-HDR (CLASSIC) REMOTE CONTROLLED RADIONUCLIDE APPLICATOR SYSTEM JAQ NUCLETRON BV M/HDR NA

Patients

Seq Age Sex Outcome Treatment
1 * Other