FDA Adverse Event Injury Summary report: N

IMPL TAPERED SCR-V SBM 3. 7MM 3.5MM 16MM

MDR report key: 16152088 · Received January 12, 2023

Report

Report Number
0002023141-2023-00146
Event Type
Injury
Date Received
January 12, 2023
Date of Event
July 17, 2022
Report Date
April 28, 2023
Manufacturer
ZIMMER DENTAL
Product Code
DZE
PMA / PMN Number
K061410
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT. PLEASE SEE ASSOCIATED REPORT: 0002023141-2023-00145. ZIMMER BIOMET COMPLAINT NUMBER (B)(4). WEIGHT UNKNOWN / NOT PROVIDED. UNIQUE DEVICE IDENTIFIER (UDI) NOT APPLICABLE. ADDITIONAL PMA/510(K) NUMBER ¿K011028/K013227. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR PERI-IMPLANTITIS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THESE EVENTS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF A MANUFACTURING OR DESIGN DEFECT THAT MIGHT LEAD TO PERI-IMPLANTITIS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NON-EXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT. PLEASE SEE ASSOCIATED REPORT: 0002023141-2023-00145-1. THIS REPORT IS BEING SUBMITTED TO RELAY CORRECTED DATA AND ADDITIONAL INFORMATION. CORRECTION: THE IMPLANT PLACEMENT DATE SHOULD BE (B)(6) 2016. THE FOLLOWING SECTIONS ARE BEING REPORTED: B4: DATE OF THIS REPORT WAS UPDATED. D6A: IMPLANT PLACEMENT DATE WAS UPDATED. G3: DATE RECEIVED BY MANUFACTURER WAS UPDATED. G6: TYPE OF REPORT WAS UPDATED. H2: TYPE OF FOLLOW UP WAS UPDATED. H10: NARRATIVE/DATA WAS UPDATED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT EXPERIENCED PAIN AND BONE LOSS AROUND THE IMPLANTS AT TOOTH LOCATIONS #3 AND #20. THE IMPLANTS WERE REMOVED. SYMPTOMS AS A RESULT OF THE EVENT: ABSCESS AND PAIN.

Description of Event or Problem · 0

NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1963568 IMPL TAPERED SCR-V SBM 3. 7MM 3.5MM 16MM DENTAL IMPLANT DZE ZIMMER DENTAL 62336680

Patients

Seq Age Sex Outcome Treatment
1 60 YR Female Required Intervention