FDA Adverse Event Malfunction Summary report: N

BD¿ PRN ADAPTER

MDR report key: 16148417 · Received January 12, 2023

Report

Report Number
3014704491-2022-00740
Event Type
Malfunction
Date Received
January 12, 2023
Date of Event
October 14, 2022
Report Date
February 10, 2023
Manufacturer
BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
Product Code
FPA
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INITIAL REPORTER PHONE #: (B)(6). A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY: A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 9050858. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. ALTHOUGH PHOTOS WERE SUBMITTED FOR EVALUATION, THE AFFECTED SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, FUNCTIONAL TESTING WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RESULTS OF THESE SHOW THAT THE TESTED UNITS PERFORMED WITHIN PRODUCT SPECIFICATIONS. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THEY FOUND BLACK SPOTS IN THE BD¿ PRN ADAPTER BEFORE USE AND UNUSED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: UNOPENED HEPARIN CAP BLACK SPOTS FOUND BEFORE USE, UNUSED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THEY FOUND BLACK SPOTS IN THE BD¿ PRN ADAPTER BEFORE USE AND UNUSED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: UNOPENED HEPARIN CAP BLACK SPOTS FOUND BEFORE USE, UNUSED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1238527 BD¿ PRN ADAPTER INTRAVASCULAR ADMINISTRATION SET FPA BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU 1313602

Patients

Seq Age Sex Outcome Treatment
1 Unknown