FDA Adverse Event Malfunction Summary report: N

OMNIPOD DASH¿ INSULIN MANAGEMENT SYSTEM

MDR report key: 16148284 · Received January 12, 2023

Report

Report Number
3004464228-2023-01061
Event Type
Malfunction
Date Received
January 12, 2023
Date of Event
November 10, 2022
Report Date
January 10, 2023
Manufacturer
INSULET CORPORATION
Product Code
LZG
UDI-DI
10385082000146
PMA / PMN Number
K192659
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

A TECHNICAL ASSESSMENT OF THE DEVICE DESIGN IDENTIFIED THE ROOT CAUSE OF THE THERMAL EVENT TO BE THE DASH PDM CHARGING VOLTAGE EXCEEDING THE BATTERY SPECIFICATION, DEFINED AS OVERCHARGING. FIELD SAFETY CORRECTIVE ACTION HAS BEEN INITIATED BY INSULET CORPORATION.

Description of Event or Problem · 0

IT WAS REPORTED BY THE PATIENT THAT THE OMNIPOD PERSONAL DIABETES MANAGER (PDM) GOT EXTREMELY HOT WHILE CHARGING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1238461 OMNIPOD DASH¿ INSULIN MANAGEMENT SYSTEM PUMP, INFUSION, INSULIN LZG INSULET CORPORATION PT-000011 7536 10385082000146

Patients

Seq Age Sex Outcome Treatment
1 11 YR Female